Gas Check; View Data - QED BIOGAS 5000 Operating Manual

Gas analyzer
Table of Contents

Advertisement

BIOGAS 5000 Gas Analyzer
7.4.2

Gas Check

This option displays the 'Gas Check' menu and enables the operator to zero
and span the gas channels on the instrument. Historical/previous gas
checks data can also be viewed and factory settings can be restored.
1) Select the 'Menu' button on the front of the analyzer to display the 'Device Settings'
menu. Press the soft key to display 'Operation Settings'.
2) Select 'Key 2 – Gas Check' and the following menu is displayed:
Figure 14—Gas Check
3) For more information about the Gas Check Menu please refer to section 9.0 – Calibration.
4) Select soft-key 'Exit' to exit operation settings and return to the main screen.
7.4.3

View data

This option enables the operator to view the readings collected and stored
on the instrument. Readings many be downloaded to the optional LANDTEC
System Gas Analyzer Manager (LSGAM) software if further analysis is
required.
1) Select the 'Menu' button on the front of the analyzer to display the 'Device Settings'
menu. Press the soft-key to display 'Operation Settings'.
2) Select 'Key 3 – View Data' and the following screen is displayed:
seconds.
5K-MNL-BIOGAS5000
Page 35