Timers - QED BIOGAS 5000 Operating Manual

Gas analyzer
Table of Contents

Advertisement

5K-MNL-BIOGAS5000
Figure 12—Operation settings menu
7.4.1

Timers

The timers function enables the operator to set standard purge times and
set auto-power off if the unit is untouched for the period of time specified.
1) Select 'Key 1 – Timers' and the following screen is displayed:
Figure 13—Timers
1)
Select 'Key 1' to edit the purge time. Enter the 'Pump Running Time' in seconds; this is
the length of time you wish to run the pump to draw the sample, e.g. key in 030 then
press the 'Enter' key to accept.
2)
Select 'Key 2' to edit the auto power off time. Enter the 'Auto power off' in minutes;
the instrument will automatically power off to preserve the battery life after the
specified time if no activity has occurred on the instrument. Press the 'Enter' key to
accept.
3)
Select the soft-key 'Exit' key to exit the screen and return to the 'Operation settings'
menu.
Note:
Page 34
Setting the purge time and auto power off functions to zero, disables the
option. It is not recommended to reduce the purge time to below 30
BIOGAS 5000 gas analyzer