READING EXAMPLE
To read the gross weight on the DAT 200 it is needed to read the addresses from 2 to 5 of the Input Area.
To read the net weightit is needed to read the addresses from 6 to 9 of the Input Area.
Whwn the display shows the gross weight value of 12351 in the corresponding bytes there will be:
Byte 2
Hex
00
WRITING EXAMPLES
To write the set-up parameters following the example:
In the byte 132 (Command Register) write value Hex 7FFF. This value opens the writing area of the
DAT 200.
Example: to change the default values of the DAT 200 like the Capacity of the load cells, the Sensitivity
and Division value to 15000, 2.9965 and 2:
Capacity
Byte 134
Hex
00
Dec
Sensitivity
Byte 138
Hex
75
Dec
29965
Division
Byte 140
Hex
00
Dec
Save the data by writing the value Hex 20 in Command Register. In case Zero and FS Calibration is
not performed, put value Hex 0 in Command Register.
N.B. The DAT 200 does not accept writing of the same values already written.
To perform Zero and FS Calibration it is not needed to abilitate the internal Writing Area of the DAT 200.
Zero Calibration:
Whit empty system put Hex 10 in Command Register (byte 132). The new Zero value is stored. To de-
activated the Zero Calibration command put value Hex 0 in Command Register.
Full Scale Calibration:
Put a know weight on the system and write its value in the Data Register (from byte 128 to 131). Put
value Hex 11 in Command Register. The weight value will be displayed.
To store the new value put value Hex 20 in Command Register.
To end the procedure put value Hex 0 in Command Register.
Byte 3
Byte 4
00
30
0
12351
Byte 135
Byte 136
00
3A
15000
Byte 139
0D
Byte 141
0D
13
Byte 5
3F
Byte 137
98
Page 39