Download Print this page

Emerson Rosemount 2120 Quick Start Manual page 17

Level switch vibrating fork
Hide thumbs Also See for Rosemount 2120:

Advertisement

March 2022
3.5.3
Relay DPCO cassette (standard version)
Figure 3-3: Relay DPCO Cassette, Standard Version (Green Label) – Code
V
1
Note
A DPST (Double Pole, Single Throw) on/off switch must be fitted for safe
disconnection of the power supply. Fit the DPST switch as near as possible to
the level switch. Keep the DPST switch free of obstructions. Label the DPST
switch to indicate it is the supply disconnection device for the level switch.
Table 3-5: Electrical Parameters
Parameter
U
I
Quick Start Guide
2
3
Fuse 0.5 (T)
DPST
N
L
0V
+V
Value
20 - 60 Vdc or 20 - 264 Vac (50/60Hz)
< 6 mA
4
5
6
7
NC
C
NO
NC
Quick Start Guide
8
9
C
NO
17

Advertisement

loading