(Please read this manual carefully before using the column to ensure performance and life.) Introduction The Shodex Asahipak NH2P series is packed with a polymeric gel containing chemically bonded polyamine groups. Under the chromatographic conditions generally employed for the silica-base amino columns, the NH2P column provides equal resolution and far greater reproducibility.
When replacing buffer solution to acetonitrile solution, also flow 100% water to the column before flowing acetonitrile. Attention! Shodex line filter, Asahipak NH2P-LF (F6710100), is a filter for removing undesired contaminants in the eluant. By installing the filter between the pump and the injector, deterioration...
Operation Manual Shodex Asahipak NH2P-40 3E Version 150305EJ Attention! Column equilibration for using acetonitrile/buffer eluent Anions in the buffer interact with the amino functional groups of NH2P packing gel. The reproducibility of the analysis will not be attained until the ionic equilibrium is attained between the anions and the amino groups.
5) When the column is not used for two weeks or more, replace the in-column solvent with the shipping solvent, remove it from the LC system, close each end with a stopper, and store it at controlled room temperature. 6) Contact Shodex website (www.shodex.de) or Shodex partners regarding product and analysis applications.
Page 6
User’s exclusive remedy and Showa Denko K.K.‘s exclusive liability for such claim are limited to the replacement of the Shodex Column in question. In no event is Showa Denko K.K. liable for any indirect, incidental or consequential damage arising out of in connection with the Shodex Column, whether or not such damage is allegedly based on breach of warranty, negligence or otherwise.
Page 7
Version 150305EJ Product Information: Shodex HILIC Columns Shodex HILIC columns are polymer-based (polyvinyl alcohol particles) for better stability (pH 2-13). This results in a longer life time of the columns than silica-based columns. They can be used in a temperature range from 4 to 60°C. The HILIC columns are suitable for RI (isocratic), evaporative light scattering, corona charged aerosol and LC-MS/MS detectors (gradient mode).