Chapter 1
Operating Elements
Manual Mode
Manual mode provides a direct printout of the real-time ECG with full control of parameter selection.
To start the manual recording of a real-time ECG, press the MANUAL Printout key
To stop the manual recording (printout) press the STOP key
The printout provides you with the following:
•
Six (selected) leads with lead identification.
•
On the lower edge, the chart speed, user identification and filter settings (if on).
•
At the top, the heart rate as current average of 4 beats, trace sensitivity, and the time and
date
The following can be freely chosen during or before the recording:
Lead Group
Note:
The LCD only displays three leads at one time. When the lead forward or lead backward
key is pressed, the following /preceding three lead group is displayed
Leads
For software versions higher than 4.10 it is also
possible to chose further leads. The desired leads
are activated as shown in the table.
Page 1.10
by means of the LEAD FORWARD and LEAD BACKWARD
key
1
2
The following lead groups are selectable:
•
I, II, III
(Cabrera: aVL, I, -aVR / II, aVF, III)
•
V1, V2, V3 /
•
II, aVF, III
/
•
V4, V5, V6
/
ALT
SCHILLER AT-2plus 6-Channel ECG Unit
SERVICE HANDBOOK Issue d May 2002
aVR, aVL, aVF
V4, V5, V6
V2, V4, V5
V7, V8, V9
Select leads
Entry Key Sequence
Lead group
Rhythm
0
II, avF, III, V2, V4, V5
Rhythm
1
II, avF, III, V2, V4, V5
Left posterior
2
V4, V5, V6, V7, V8, V9
Left posterior
3
V4, V5, V6, V7, V8, V9
Right precordial up to V5r
4
V1, V2, V3, V3r, V4r, V5r
7
9
Right precordial up to V5r
5
V1, V2, V3, V3r, V4r, V5r
Right precordial up to V6r
6
V1, V2, V3r, V4r, V5r, V6r
Right precordial up to V6r
7
V1, V2, V3r, V4r, V5r, V6r
NEHB
8
D, A, J
NEHB
9
D, A, J
Confirm
ON
OFF
ON
OFF
ON
Press STOP
key
OFF
ON
OFF
ON
OFF