Siemens SITRANS F C MC2 Operating Instructions Manual page 28

Coriolis flowmeters
Table of Contents

Advertisement

Commissioning
6.2 Zero point adjustment
MASS 6000
Choose menu entry "Reset mode"
-> Zero adjust
-> Zero adjust auto
For further information on performing an auto zero point adjustment please refer to the
relevant transmitter Operating Instructions.
● After count down (30 s.), the actual zero point is displayed and the meter ready for
26
operation.
Figure 6-1
Best practise zero adjustment with a by-pass line and two shut-off devices
SIFLOW FC070
Choose the SIMATIC PDM menu
-> Device
-> zero adjust
Operating Instructions, 10/2009, A5E02154544-02
SITRANS F C MC2