10.5.8
SIGNATURE
Sub-Menu
MODULE
Menu Item
UPDATE SIG.MOD?
Menu Item
DEFAULT SIG.MOD?
Menu Item
UPDATE DAC SETUP
page 74
008-0608-00
Updates the information stored in the transducer's Signature Module with any new
settings that is entered into the instrument with the SETUP menus. The settings that
will be updated are:
EXCITATION
FULL SCALE MV/V
FULL SCALE VALUE
SHUNT CAL VALUE
DISPLAY. UNITS
Note that changes made to the instruments display setup (such as decimal point and
count-by settings) are not updated into the Signature Module.
Updating enters these new settings from the instrument into the Signature Module. If
you do not update the Signature Module, connect a different transducer to the instru-
ment and then re-connect the original transducer again, the instrument will revert to
the settings that are stored inside the Signature Module.
The options for this menu item are:
"
" will not update the Signature Module.
NO
"
" will display
YES
WORKING
This menu item will overwrite any user settings in the Signature Module with the origi-
nal default settings. The Signature Module always maintains a copy of the original
default settings that were created at the factory.
The options for this menu item are:
"
" will not restore original default settings.
NO
"
" will display
YES
WORKING
tings, then display
DONE
This menu item determines whether or not the
log Output settings are overwritten with information from a Signature Module when a
new transducer is detected by the instrument.
The options for this menu item are:
"
" will allow a newly detected Signature Module to update the Analog Output
YES
settings.
"
" will not allow automatic updating of the Analog Output settings.
NO
, update the Signature Module, then display
, update the Signature Module, with the original factory set-
.
DAC. FULL SCALE
.
DONE
and
Ana-
DAC. ZERO SCALE