This menu item performs a calibration according to what was entered in the
and
BRATION TYPE
CALIBRATION DATA
NOTICE
Before performing a calibration,
the transducer must be connected to the instrument,
the channel's Excitation and Signal Jumpers must be set,
the
must be selected (see "CALIBRATION TYPE Menu Item"
CALIBRATION TYPE
on page 113), and
the
must be entered (see "CALIBRATION DATA Sub-Menu"
CALIBRATION DATA
on page 115).
NOTICE
For maximum accuracy, allow at least twenty minutes of warm-up with the exci-
tation voltage applied to the transducer before calibration.
If the
is Shunt Calibration...
CALIBRATION TYPE
The display will read
DOING SHUNT CAL
00000 and UNIT are previously entered as
When you have applied this load to the transducer, press [ENTER].
The display will then read
pletion, the display will indicate
mode.
If the
is 2-Point Known Load Calibration...
CALIBRATION TYPE
The display will read
DOING 2POINT CAL
(where 00000 and UNIT are previously entered as
). When you have applied this load to the transducer, press [ENTER].
UNITS
The display will read
WORKING
previously entered as
applied this load to the transducer, press [ENTER].
Upon completion, the display will indicate
the RUN mode.
If the
is 3-Point Known Load Calibration...
CALIBRATION TYPE
The display will read
DOING 3POINT CAL
(where 00000 and UNIT are previously entered as
). When you have applied this load to the transducer, press [ENTER].
UNITS
The display will read
WORKING
previously entered as
applied this load to the transducer, press [ENTER].
The display will read
WORKING
previously entered as
applied this load to the transducer, press [ENTER].
Upon completion, the display will indicate
the RUN mode.
menu items.
, and prompt you to
ZERO SCALE VALUE
and calibration will take place. Upon com-
WORKING
and the instrument will return to the RUN
DONE
, and prompt you to
, then
APPLY 01000 UNIT
and
KNOWN POINT 2/2
DISPLAY. UNITS
and the instrument will return to
DONE
, and prompt you to
, then
APPLY 00500 UNIT
and
KNOWN POINT 2/3
DISPLAY. UNITS
, then
APPLY 01000 UNIT
and
KNOWN POINT 3/3
DISPLAY. UNITS
and the instrument will return to
DONE
SC Series Instruction Manual
CALI-
(where
APPLY 00000 UNIT
and
DISPLAY. UNITS
APPLY 00000 UNIT
and
KNOWN POINT 1/2
DISPLAY.
(where 01000 and UNIT are
). When you have
APPLY 00000 UNIT
and
KNOWN POINT 1/3
DISPLAY.
(where 00500 and UNIT are
). When you have
(where 01000 and UNIT are
). When you have