Internal Arrangement - Honeywell SC1000 Owner's Manual

Sensotec sensors signal conditioning, self-calibrating digital indicators
Table of Contents

Advertisement

Removal of rear panel
Remove of case from entire circuit
board assembly

4.3.8 Internal Arrangement

CAUTION
Use a #0 Phillips screwdriver on the black screws to avoid damaging them.
Failure to comply with these instructions may result in product damage.
The rear panel must be removed in order to install or remove channels.
Step 1: Remove the four black Phillips-head machine screws that secure the back
panel of the instrument to the case. These screws are located on the rear of
the case, one at each of the four corners.
Step 2: Remove the cable shield connection screw to allow the installation or
remove of channels per "Input or Output Channel Installation Procedure" on
page 42.
Step 1: Remove the two Phillips-head screws that secure the front panel.
Step 2: Remove the front panel and disconnect its two connecting cables.
Step 3: Remove the four black Phillips-head machine screws that secure the back
panel of the instrument to the case. These screws are located on the rear of
the case, one at each of the four corners. Do not remove the cable shield
connection screw in the center.
Step 4: The circuit boards will slide out of the rear of the case as a unit.
The figure below shows the names and locations of the printed circuit boards.
Front Panel/Display Board
Assembly
SensoBus Backplane
Board
Figure 4-3: Internal Arrangement of SC1000 and SC2000
Below is a description of each printed circuit board.
The SensoBus Backplane Board serves as the connection between all
Microprocessor
Board
Up to 4 Input or Output
Channel boards
(1 shown)
Rear Panel
SC Series Instruction Manual
Power Supply
Board
page 29

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sc3004Sc2000Sc2001

Table of Contents