Dura Mag ™ Battery Powered Electromagnetic Flow Meter Installation, Operation and Maintenance Manual for use with Procomm GO Converter 30122-53 Rev. 2.0 April 25, 2022...
Safety Warnings When installing, operating, and maintaining McCrometer equipment where hazards may be present, you must protect yourself by wearing Personal Protective Equipment (PPE) and be trained to enter confined spaces. Examples of confined spaces are manholes, pumping stations, pipelines, pits, septic tanks, sewage digesters, vaults, degreasers, storage tanks, boilers, and furnaces.
When uncrating the Dura Mag, any damage due to rough or improper handling should be reported to the transportation firm and McCrometer. If for any reason it is determined that the unit or parts of the unit should be returned to the factory, please contact McCrometer for clearance prior to shipment.
Preparing for a New Installation 3 .2 Pipe Run Requirements Flow meters are velocity sensing devices and are vulnerable to certain upstream disturbances. Because of this, meters need certain lengths of straight pipe before and after the meter. These distances relate to the diameter of the pipe used.
The flow of the medium should correspond to the direction shown by the arrow on the sensor. NOTE 4 .1 Wafer Style Meter Installation Install the Dura Mag between two flanged end pipes. Shown are customer’s Hardware and gaskets are provided, but customers must pipe ends and flanges supply the flanged pipe ends.
Connections to the sensor must be made with cable supplied by McCrometer specifically for that purpose. Do not substitute the supplied cable with other types of cable, even for short runs. For repairs or added lengths of cable, the entire cable between the sensor and the converter must be replaced.
Converter Wiring and Connection 6 .0 CONVERTER WIRING AND CONNECTION 6 .1 External Output Connectors 10-35VDC Power/4-20mA Output And Pulse Output The flow meter has one dedicated port and two optional ports on the back side of the electronics enclosure. The dedicated port (center) is used to download data logger information or for meter service and troubleshooting.
Converter Wiring and Connection 6 .4 Wiring Diagrams TERMINAL BLOCK ASSIGNMENTS ATTACHES TO Coils Harness BLACK CHASSIS LUG Terminal Port Wire Color YELLOW Green/Yellow Yellow Chassis Lug Black GREEN / YELLOW Electrodes Harness Terminal Port Wire Color BLUE BLACK Green/Yellow PINK Pink GREEN / YELLOW...
(Figure 14) • Signals and associated wire colors in the McCrometer SmartOutput™ interface cable are identified together in the top row of the table at right. • Corresponding wire colors for transceivers from each compatible AMI vendor are identified in the columns under the top row.
Grounding for meters with minimal ground noise 1 . Preferred method of grounding Dura Mag meters come standard with a set of grounding rings for use with the preferred method of grounding Dura Mag meters. This method can be used for all installations, but it is required for non-conductive or internally coated pipe. When pipes are non-conductive, such as PVC or internally coated pipe, you must substitute direct grounding with grounding rings.
Page 13
(Figure 16) Next, connect the two mating pipe sections with a 12 gauge wire or equivalent and connect them both to the earth ground. The Dura Mag should be electrically isolated from the pipeline. 12 gauge ground wire...
General The flow meter comes pre-configured from the factory based on the installation parameters provided to McCrometer at the time of order. Other than activating the display, there is nothing required of the user for the basic operation of the flow meter.
This is only a summary of the configuration tool’s functionality. The software has complete set-up and operation instructions included. The software can be downloaded from the McCrometer Web site. Figure 22 . Plug in mini-USB cable Software Operation Remove the cap that protects the USB port and plug in a mini-USB cable.
ProComm GO Converter 3000 This procedure applies to Dura Mag and McMag3000 ow meters with the ProComm GO converter. It describes how to install batteries in a new meter and how to replace batteries using the replacement battery kit.
Page 17
ProComm GO Battery Installation and Replacement Battery Removal and Replacement II. Removing the batteries Unplug tall connectors to sensor, outputs, and Loosen the captive screws opposite of the hinge and power. lift the battery cover. Hinge Hinge screws Connectors Unplug the batteries. Cut the two zip ties securing the batteries.
Page 18
ProComm GO Battery Installation and Replacement Battery Removal and Replacement III. Installing the batteries and restoring the power Cut the zip tie holding the 10. Pass the new zip tie through the 11. Wrap the new zip tie around the dry pack to the hinge on the slots in the internal hinge.
Gallons per hour * All Dura Mag meters are calibrated in a NIST traceable gravimetric test stand with a minimum straight run of 10D upstream and 2D downstream. A calibration certi cation report is provided with each Dura Mag meter certifying the speci cation accuracy in our laboratory test.
Page 21
Lifetime guarantee METER GROUNDING RECOMMENDATIONS Dura Mag meters come standard with a set of grounding rings for use with the preferred method of grounding Dura Mag meters. Grounding the meter body for safety according to national (NEC) or local electrical codes is recommended on ALL meter installations.
19.0 13.5 Pipe size Pipe size Dura Mag S Series and P Series meters are available in lay lengths compatible with products from other meter suppliers. See the tables below for matched lay length options. Other Supplier S Series Other supplier...
Page 23
Speci cation Sheet Dimensions and Weights ™ SPECIFICATIONS ProComm Dura Mag Flow Meter 9.0 SPECIFICATIONS PROCOMM GO CONVERTER SPECIFICATIONS Physical Speci cations Electronic Housing Diecast aluminum, powder coated enclosure w/ tamper resistant seal, 6½” x 6½” x 43/8” tall Battery: Standard: three 3.6V lithium-thionyl chloride (Li-SOCI2) D size...
• Call the McCrometer Customer Service Department and ask for a Return Authorization (RA) number. • Ship the meter in the original packaging, if possible. Do not ship manuals, power cords, or other parts with your unit unless required for repair.
The batteries and their con guration are speci c TO EQUIPMENT AND PRODUCTS ARE EXCLUSIVE AND to the Dura Mag ow meter and only McCrometer batteries IN LIEU OF ANY AND ALL OTHER WARRANTIES OF can be used. Any use of batteries other than those supplied QUALITY OR PERFORMANCE, EXPRESS, IMPLIED OR from McCrometer will void the above Warranty.