Table 5-17: Calibration States - AMC -UT M-400 Series User Manual

Modbus rtu sensor/transmitter
Table of Contents

Advertisement

AMC-UTx-M-xx-400 Series Digital Transmitter
Menu Item
Buffering
Find Zero
Apply Gas
Find Gas
Find Span
When the calibration procedure complete its cycle, the result will either be a PASS or FAIL as
shown by the following two images.
ADAPTiCal
- - - - - - - - - - -
Cal Passed
OK
ADAPTiCal
- - - - - - - - - - -
Cal Failed
OK
In either case, the OK command is highlighted, therefore, press the ENTER button to return back
to the sensor's service menu.

Table 5-17: Calibration States

Description
An internal buffer needs to be completely
filled with gas reading before the gas level can
be evaluated
The gas concentration for zero gas is
determined
Procedure is ready for the user to apply gas.
Calibration routine will wait until it has
detected the application of gas.
Calibration is underway with the specified gas
applied. This state will remain active until it
either passes or fails.
Displayed when the calibration procedure is successful
OK command is highlighted
Displayed when the calibration procedure has failed
OK command is highlighted