AMC -UT M-400 Series User Manual page 52

Modbus rtu sensor/transmitter
Table of Contents

Advertisement

AMC-UTx-M-xx-400 Series Digital Transmitter
With 'ADAPTiCal' highlighted (as above), press
selected sensor. Once initiated, the user will be presented with the following screens in which
the background and span level concentrations used for the calibration can be entered:
Once the background and span level gas concentrations are entered, the calibration procedure
will start automatically and the state of the calibration will be displayed on the display as shown
below. Throughout the entire calibration procedure, the Abort command remains highlighted
therefore the procedure can be aborted at any time by pressing the
ADAPTiCal
- - - - - - - - - - -
Buffering
Abort
During calibration, the procedure will transition through the following set of states:
ENTER
to initiate a calibration sequence for the
Current state of the calibration procedure
Abort command remains highlighted throughout the entire
procedure
52
ENTER
button.