Fluke 5820A Operator's Manual page 220

Oscilloscope calibrator
Hide thumbs Also See for 5820A:
Table of Contents

Advertisement

5820A
Operators Manual
Calibrator
Mainframe Freq.
(MHz)
1,525,000,000
1,575,000,000
1,650,000,000
1,750,000,000
1,850,000,000
1,925,000,000
1,975,000,000
2,025,000,000
2,075,000,000
1.2 Vpp
625,000,000
725,000,000
900,000,000
1,125,000,000
1,250,000,000
1,325,000,000
1,375,000,000
1,425,000,000
1,475,000,000
1,525,000,000
1,575,000,000
1,650,000,000
1,750,000,000
1,850,000,000
1,925,000,000
1,975,000,000
2,025,000,000
2,075,000,000
0.800 Vpp
625,000,000
725,000,000
900,000,000
8-8
Table 8-4. Leveled Sinewave Verification: Flatness (cont.)
Measured
10 MHz Power Out
Power Out (w)
Deviation
(w)
(w)
Calibrator
Mainframe
Flatness
% Deviation
Spec. (%)
± 5.0
± 5.0
± 6.0
± 6.0
± 6.0
± 6.0
± 6.0
± 6.0
± 6.0
± 5.0
± 5.0
± 5.0
± 5.0
± 5.0
± 5.0
± 5.0
± 5.0
± 5.0
± 5.0
± 5.0
± 6.0
± 6.0
± 6.0
± 6.0
± 6.0
± 6.0
± 6.0
± 5.0
± 5.0
± 5.0

Advertisement

Table of Contents
loading

Table of Contents