Edge Verification: Amplitude; Edge Verification: Frequency; Edge Verification: Duty Cycle - Fluke 5820A Operator's Manual

Oscilloscope calibrator
Hide thumbs Also See for 5820A:
Table of Contents

Advertisement

7-17. Edge Verification: Amplitude

Nominal Value
Frequency
(V p-p)
(Hz)
0.005
1 kHz
0.005
10 kHz
0.005
100 kHz
0.01
100 kHz
0.025
100 kHz
0.05
100 kHz
0.1
100 kHz
0.25
100 kHz
0.5
100 kHz
1
100 kHz
2.5
100 kHz
2.5
10 kHz
2.5
1 kHz

7-18. Edge Verification: Frequency

Nominal Value
(V p-p)
Frequency
2.5
1 kHz
2.5
10 kHz
2.5
100 kHz
2.5
1 MHz
2.5
10 MHz

7-19. Edge Verification: Duty Cycle

Nominal Value
(V p-p)
Frequency
2.5
1 MHz
Table 7-13. Edge Verification: Amplitude
Measured Value
(V p-p)
Table 7-14. Edge Verification: Frequency
Measured Value
(Hz)
Table 7-15. Edge Verification: Duty Cycle
Measured Value
(%)
Deviation
1-Year Spec.
(V p-p)
0.0003 (typical)
0.0003 (typical)
0.0003 (typical)
0.0004 (typical)
0.0007 (typical)
0.0012 (typical)
0.0022 (typical)
0.0052
0.0102
0.0202
0.0502
0.0502
0.0502
Deviation
(Hz)
Deviation
1-Year Spec.
(from 50 %)
Maintenance
Verification Tables
(V p-p)
1-Year Spec.
(Hz)
330E-6
3.3E-3
33E-3
330E-3
3.3E-0
(%)
5 (typical)
7
7-19

Advertisement

Table of Contents
loading