1 Click on the GC parameters icon or select Instrument > GC parameters.
2 Select Events to display the Runtime Events table.Enter the Time when the reference valve
3 Select the Valve number assigned to the reference valve for the Position, and select On for
Figure 39.
Figure 40.
4 Enter the Time when the reference valve closes, select Valve for the Event Type, select the
5 Click OK to close the GC Edit Parameters window and add this timed event to the method.

Verify the correct column is installed

1 In the MassHunter GC parameters window select Configuration > Columns
2 Verify that the following column is installed and selected for this method. If this column is
3 Save the method.
G7050A - Reference Compound Introduction Valve
opens, select Valve for the Event Type, See
the Setpoint.
Adding the valve operation event to the Runtime Events table
Adding the valve operation event to the Runtime Events table (continued)
valve number assigned to the reference valve for the Position, and select Off for the
Setpoint.
not installed you must install it first and return here to verify and save the method.
For 8890/7890 Agilent DB-5, 60 m ? ~ 0.25 mm, 0.1 um (p/n 122-5061)
For 9000 GC use the equivalent Agilent DB-5, 60 m x 0.25 mm, 0.1 um column
Figure 39
on page 31.
31