Intended Use; Operational Limitations; Preparation For Use; Visual Inspection - NOVOTEST TP-2020 Operating Manual

Coating thickness gauge
Table of Contents

Advertisement

Operation manual
Coating Thickness Gauge NOVOTEST TP-2020

2.1 Operational limitations

The device should be operated under the influence of factors and taking into account the parameters
of the test object under the specified technical characteristics, and the device must be used within its
specification.
The device shall be operated only by a qualified user familiar with the operation manual.
After transporting the device to the place of operation at a negative ambient temperature and placing
it into a room with a positive temperature, keep the device in its packaging for at least 6 hours to avoid
failure due to the moisture buildup.

2.2 Preparation for use

2.2.1 Visual inspection

Inspect the device visually and make sure that there is no mechanical damage to the processing unit
(4), probe (10), connector (7), and connecting cable (9).

2.2.2 Battery installation

Place a battery into the battery compartment (5) after pushing and pulling down the cover of the
battery compartment until detaches. The power element or batteries should be placed in line with the
polarity indicated on the device. Close the battery compartment cover until it snaps into place.

2.2.3 Probe connection

Connect the used probe (10) by plugging the connection cable (9) into the probe socket (7) on the
processing unit (4) so that the red dots on the plug and socket are aligned (fig. 2.1).
Note – The probe can be connected and disconnected from the electronic unit only with the device
powered off.
Warning!
To prevent damage to connectors and cables, follow the instructions for handling those connectors
below!
The connectors used in the device (fig. 2.2) consist of two parts: socket and plug (pin) of the cable.
Page 11

2 INTENDED USE

Figure 2.1 – The probe connection
RDC.OD.TP-2020.000 OM

Advertisement

Table of Contents
loading

Table of Contents