Guidance and manufacturer's declaration - electromagnetic immunity
Model POC-W181 is intended for use in the electromagnetic environment specified below. The
customer or the user of the model POC-W181 should assure that it is used in such an environ-
ment.
IEC 60601 test
Immunity test
level
3 Vrms
Conducted RF
150 kHz to 80
IEC 61000-4-6
MHz
3 V/m
Radiated RF
80 MHz to 2,5
IEC 61000-4-3
GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) tele-
phones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the mea-
sured field strength in the location in which the model POC-W181 is used exceeds the applica-
ble RF compliance level above, the model POC-W181 should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such
as reorienting or relocating the model POC-W181.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Compliance
Electromagnetic environment - guid-
level
ance
Portable and mobile RF communica-
tions equipment should be used no
closer to any part of the model POC-
W181, including cables, than the recom-
mended separation distance calculated
from the equation applicable to the fre-
quency of the transmitter.
Recommended separation distance
d = 1,2
d = 1,2
d = 2,3
3 Vrms
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufac-
turer and d is the recommended separa-
3 V/m
tion distance in meters (m).
Field strengths from fixed RF transmit-
ters, as determined by an electromag-
netic site survey,
the compliance level in each frequency
range.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
ix
80 MHz to 800 MHz
800 MHz to 2,5 GHz
a
should be less than
b
POC-W181 User Manual