Calibration station
Hide thumbs Also See for GT Series:

Advertisement

Fig. 3 J25 Connector
Calibration Station Gas Supply:
The gases can be plumbed in by using nylon tubing and
connect these to swage connector gas inlets at the top of
the calibration station. Outer diameter is 6mm. (
1. Air
2. PPM
3. LEL
4. CO (if applicable)
Swagelok connectors must be finger tightened then,
using two spanners, tighten a further 1
not overtighten.
A gas pressure of 590 to 690 mBar (8.5 to 10 PSI)
should be provided to these gas inlets.
Calibration station computer connection and driver software:
* The calibration station has a USB connector that can be easily connected to a free USB port on the
computer. (*not applicable to 6761 1 and 6761 1Q)
Install the software from the enclosed CD-ROM to initially set up the gases.
Start the software. The first time you will be asked to select the com-port.
Install instrument in the Calibration Station
Make sure that the power contacts, on the back of the GT instrument and on the calibration station (shown
in Fig. 4), are clean and free of dirt / grease. As detailed on calibration station instructions label, carefully
position the GT in its docking location, as shown in Fig.1, with the bottom part first, placing the extension
pole recess at the bottom of the GT so that it docks with the holding clip shown in Fig. 4. Push the instrument
backwards so that it is located between the two holding arms then engage the securing strap. Locate the
probe as shown in Fig. 1.
Follow the further instructions on the front of the calibration station.
Inchinnan Business Park,
Renfrew, PA4 9RG
Scotland, UK
Tel: +44 (0) 141-812 321 1
e-mail:
gmi_sales@teledyne.com
As an ISO9001 approved company, Teledyne GMI Ltd's quality assurance programmes demand the continuous assessment and improvement of
all Teledyne GMI products. Information in this leaflet could thus change without notification and does not constitute a product specification. Please
contact Teledyne GMI or their representative if you require more details.
1
/
in.).
4
5. H
S (if applicable)
2
6. - Empty -
7. - Empty -
8. Vol. Gas
1
/
turns. Do
4
www.TeledyneGFD.com
Copyright © Teledyne Gas Measurement Instruments Ltd 2021
Part Number 67295 Issue 6 (10/09/21)
Calibration Station
Fig. 4 Instrument Location
Service & Calibration Division:
25 Cochran Close, Crownhill,
Milton Keynes, MK8 OAJ
Tel: +44 (0) 1908 568867
e-mail:
gmi_service@teledyne.com

Advertisement

loading

This manual is also suitable for:

6710967109q