Fig 2-9 Store (Edit) Mode Operationo Flow-Chart - Prodigit 33501 Series Operation Manual

High power electronic load
Table of Contents

Advertisement

2-8 PRODIGIT
Press Exit key

Fig 2-9 STORE (EDIT) MODE OPERATIONO FLOW-CHART

2-14-2 TEST MODE
The TEST Mode of Auto-Sequence function is entered by press S3 + S4 key
simultaneously, the S3 and S4 LED are ON to indicate Auto-sequence mode, then
the Test mode of Auto-sequence function is proceed by press START key.
The TEST MODE flow chart is described below:
2-14-2-1
2-14-2-2
2-14-2-3
2-14-2-4
S3+S4
Show n1-n9
Exit
Press Exit key
After press START key, the 33501 series controls within the mainframe to
recall correspond memory which had been stored in Auto-sequence
(n1~n9) memory.
The sequence start from (Step 0 - t1 - t2), then (step 1 - t1 - t2), and so on
until last step or stop by press EXIT key.
The two digit LED display will show GO (flash) if all the test in all module
is pass, and will show nG (flash) if there is at least one failure during the
test.
User can press Start key to continue another test, or the 33501 series can
quit from Auto-Sequence mode by press EXIT key.
Initial=n1~n9
Press Store
Store Mode
Show SP= 0
Press Up/Down key to set Bank No.
Set Bank
Set State No.1~5
Set State
Press Store key to next item
When press Store key, LED show t1
Show t1
Press Up/Down key to set T1,
Set T1
t1 = 01-99, Unit = 100mS
Press Store key to next item
When press Store key, LED show t2
Show t2
Press Up/Down key to set T2,
Set T2
t1 = 00-99, Unit = 100mS
Press Store key to next item
Yes
Step = F?
Is last step (16 step)?
Yes
End?
Sp=Sp+1
1.Press S3+S4 key
2.Show AUTO SEQ. No.1-No.9,
Press Up/Down key to set n1-n9
3.Press Store key to Store Mode
Sp = 0 ~ F(16 step)