2-14 AUTO SEQUENCE testing function description

There are two modes in AUTO SEQUENCE function, EDIT MODE and TEST MODE, The
AUTO SEQ mode can be entered by press S3 + S4 key simultaneously, then press STORE
key again to enter the EDIT MODE, or press START key to enter the TEST MODE, Please
refer to the flow chart operation below:
2-14-1 EDIT MODE
The TEST Mode of Auto-Sequence function is entered by press S3 + S4 key
simultaneously, The EDIT mode of Auto Sequence function is entered by press S3 +
S4 key simultaneously, the S3 and S4 LED are ON to indicate Auto-sequence mode,
then the Edit mode of Auto-sequence function is proceed by press STORE key.
The EDIT MODE flow chart is described below:
2-14-1-1
2-14-1-2
2-14-1-3
2-14-1-4
Press S3+S4 key
AUTO SEQ.
Press STORE key
STORE MODE
Press Exit key

Fig 2-8 AUTO SEQ function operation flow chart

There are nine Auto Sequence (n1-n9) can be edit within 33501 series.
Each Auto Sequence has up to 16 Test step, where each step is any one
memory of 150 sets Store memory which is 30 memory Bank with 5
memory state.
Each test step has t1 (test time) and t2 (delay time), the unit is 100mS,
the range is 0.1S - 9.9S in 100mS resolution. 33501 series mainframe will
check each module GO/NG at the end of t1 (test time), the next step will
be started after duration t2 (delay time).
The test step sequence can be up to 16 step, and can be terminated by
press EXIT key if less than 16 step is required.
33501 Series Operation Manual 2-7
Display 3300C Firmware version r1-r9
Show r1-r9
Normal Mode
bn
Display bank
Display bank no.
Show n1~n9
Press STATE key
STATE MODE

Advertisement