Stryker 1188 HD Troubleshooting Manual

Stryker 1188 HD Troubleshooting Manual

Camera head and coupler
Hide thumbs Also See for 1188 HD:

Advertisement

Quick Links

ITEM:
Stryker 1188 HD Camera Head and Coupler
Increase/Decrease Brightness
Record/Picture Switch
Unscrew soaking cap of the camera cord and locate the Blue Arrow on camera
cord. Plug the camera cord into the Camera Head with the blue arrow facing Up.
Common Issues and Troubleshooting
Blurry Image
Make sure the camera is focused, and wipe the scope end and coupler window clean. Ensure
the coupler is clean and attached securely to the camera as moisture may be present from
sterilization. Re-White Balance the camera if needed.
Excess Glare
Decrease the light source power on the console. Press the Down arrow on the camera head to
decrease the shutter as needed.
1188 HD Camera Head
TroubleShooting Guide
Zoom/White Balance Adjust
Activate light source, press and Hold W Button to white balance while pointed at gauze. Hold until
White Balance Complete shows on monitor upper corner. A Quick W Press will cycle through the
Zoom.
Camera Coupler
Attach scope and light cord to camera. Ensure coupler has latch at 12 O'Clock
position, rotate clockwise if needed.
Focus Knob
Rotate Focus Knob to focus and set depth after insertion. Press Up and Down
arrows to increase or decrease Brightness.
Low Light
Increase the light source power on the console. Press the Up arrow on the camera head to
increase the shutter as needed. Check light cables for broken fibers and replace if several black
spots are present when holding cable end up to a light.
Grainy Image
Ensure the Scope end and the Coupler attachment are clean of debris. Re-White Balance the
camera and try again. Make sure the camera is focused and Decrease the light source Bright-
ness and Zoom if needed. Make sure the light cables are in good condition.
For more help or support information, please visit us at aamedicalstore.com

Advertisement

loading

Summary of Contents for Stryker 1188 HD

  • Page 1 ITEM: TroubleShooting Guide Stryker 1188 HD Camera Head and Coupler Zoom/White Balance Adjust Activate light source, press and Hold W Button to white balance while pointed at gauze. Hold until Increase/Decrease Brightness White Balance Complete shows on monitor upper corner. A Quick W Press will cycle through the Zoom.