Delta DE212 Installation And Operation Manual page 22

Table of Contents

Advertisement

The battery replacement operation must always take place outside Ex-area.
When the battery module has to be replaced, this procedure needs to be followed strictly:
Verify that the old battery module 02166 is replaced by 02465.
Battery module 02465 must be replaced by 02465 only.
Battery types other than specified must NEVER be used.
Open the bottom end of the transmitter, by pressing the two lock-sections in.
Pull the battery module gently out of the housing without jamming the O-ring.
Disconnect the battery from connector marked "P3" on the PCB, by lifting the
connector up. See section 5.3.2
Put a thin layer of non-corrosive, acid-free Vaseline on the new battery module O-ring
Connect the new battery, and push the battery module back into the housing.
Carefully make sure the O-ring and connecting wires are not jammed during this
operation.
When the operation is finished, charge the battery for 3 hours, and test all functions.
See section 3.2.5
Defective batteries must be disposed at a proper recycling facility.
Please refer to chapter 10 for further information.
D
RC AS
ELTA
WARNING:
Page 21 of 33
Document no: 03039

Advertisement