Insertion-Type Electromagnetic Probe Flowmeter
AquaProbe 2

5 Start Up And Operation

Warning.
The lithium battery used in this device may present a risk of fire or chemical burn if mistreated. Do
not recharge, disassemble, heat above 100°C or incinerate.
Replace battery with ABB Limited Part No. WABC2001 only. Use of another battery may present
a risk of fire or explosion.
Dispose of used battery promptly. Keep away from children.
Dispose of used batteries in accordance with your local regulations.
Where possible, recycle used batteries.
Contact your local environmental authority for further information regarding disposal or recycling
schemes for used batteries.
Note.
If fitting a battery on an external powered (a.c. or d.c.) unit, ensure the unit is powered during this
operation.
Each battery must be connected to the cable from the same side of the termination area as the
battery position in the lid.

5.1 Fitting the Battery

If the AquaMaster has been supplied with one or two batteries, but not connected, then proceed as
follows:
1. Remove the top cover of the transmitter – see section 3.1.3, page 19.
2. Invert the cover.
3. Slide out the connector from behind the battery retaining clamp.
4. Connect the battery or batteries to the wire connector(s) inside the top of the transmitter unit; left
hand battery to left hand connector and right hand battery to right hand connector.
5. Ensure that the end of the battery with the connection wires is pushed up against the inside end of
the top cover.
6. Push the connection centrally behind the battery retaining clamp to secure the battery.
7. Fit the top cover to the transmitter and ensure the screws are tightened fully.

5.2 Start-up

If the AquaMaster is received with a protective plastic film over the display window, remove this film before
commencing normal operation.
When the power is connected or the plastic film is removed, the AquaMaster performs a self test operation,
and indicates a successful completion with 'EE Pass' displayed.
If the display shows 'EE Fail 1', remove all power, check the sensor wiring and apply power.
If the display shows 'EE Fail 2 or 3', contact ABB.
32
5 Start Up And Operation
IM/AP
Issue 9