ABB AquaProbe 2 Installation Manual page 22

Insertion-type electromagnetic probe flowmeter
Table of Contents

Advertisement

Insertion-Type Electromagnetic Probe Flowmeter
AquaProbe 2
Caution. (Remote versions)
Make connections only as shown.
Remove foil screens
Twist the three screen wires together and sleeve them.
Twist cable pairs together
Maintain Environmental Protection at all times.
Conduit connections must provide cable entry sealing.
AquaMaster Transmitter
1
Fig. 3.5 Sensor Cable Connections (Gland/Conduit, Remote version)
Fig. 3.6 Sensor Cable Connections (Connector, Remote version)
Note. This arrangement is an option.
20
7
AquaMaster Transmitter
1
7
3 Electrical Installation
CABLE
Belden 8777
ABB Limited
1 Black
2 Green
3 Black
4 Red
5 Sleeved Ground
5 Sleeved Ground
6 Black
7 White
1 Brown
2 Red
3 Orange
4 Yellow
6 Blue
7 Violet
IM/AP
Issue 9

Advertisement

Table of Contents
loading

Table of Contents