Page 1
Flow Model CTF878 Clamp-On Tag Flowmeter Service Manual 910-254S Rev. A panametrics.com AUG 2021...
Page 3
Flow Model CTF878 Clamp-On Tag Flowmeter Service Manual 910-254S Rev. A AUG 2021 panametrics.com Copyright 2021 Baker Hughes company. This material contains one or more registered trademarks of Baker Hughes Company and its subsidiaries in one or more countries. All third-party product and company names are trademarks of their respective...
Page 5
Infrastructure Sensing, Inc. will issue a RETURN AUTHORIZATION number (RA), and shipping instructions for the return of the instrument to a service center will be provided. 2. If Panametrics instructs you to send your instrument to a service center, it must be shipped prepaid to the authorized repair station indicated in the shipping instructions.
Page 6
Warranty [no content intended for this page] Model CTF878 Service Manual...
• diagnose setup problems 1.1.2 Entering the Service Menu To enter the Service Menu, press the [MENU] key at the lower right of the CTF878 keypad. The Menu Bar replaces the Status Bar at the top of the screen. Press the [ ] arrow key to scroll from the Site Menu to the Service Menu.
19.25 Corr SNR Cancel Figure 2: Diagnostics Tests Window Press [F2] (Cancel) or [F3] (OK) to close the window and return to Operate Mode. Note: For an explanation of diagnostic parameters, refer to Chapter 4, Troubleshooting. Model CTF878 Service Manual...
Reverse Flow, and press the [ ] arrow key to enter the next prompt, or Press [ENTER] again to deselect Reverse Flow, and press the [ ] arrow key to enter the next prompt. Model CTF878 Service Manual...
Chapter 1. he next prompt asks for the Zero Cutoff. Near “zero” flow, the CTF878 may have fluctuating readings due to insufficient flow information. The zero cutoff causes velocity measurements less than the cutoff to be reported as zero. To set the cutoff:...
Page 13
Figure 4: Signal Error Limits Option The first prompt asks for the minimum error limit for the correlation signal-to-noise ratio (SNR) calculated by the CTF878. The default value is 10. The E8: SNR error message appears if the SNR falls below the limit programmed here.
The Test Option Within the Service Menu, the Test option includes three tests to ensure that the CTF878 is performing properly: Test Screen, Test Keys and Watchdog Test. To enter this option, scroll to the Test entry on the Service Menu and press [ENTER].
Page 15
Testing the Screen To test the proper functioning of the CTF878 screen, scroll to the Test Screen option in the Test Menu and press [ENTER]. The screen then shows the message, “Press Any Key To Continue.” Once you press a key, a screen with a checkerboard pattern appears, as in Figure 6 below.
Page 16
Test Keys MENU Exit Figure 7: The Test Keys Window Pressing the [F3] key returns the meter to the Operate Mode. If any key does not appear on the screen, contact the factory. Model CTF878 Service Manual...
Page 17
Figure 8: Watchdog Test Screen Press [F2] (No) to cancel the test and return to the Menu screen, or press [F3] (Yes) to start the test. The CTF878 should go blank for a few seconds, and then restart. If it does not follow this sequence, consult the factory.
Page 18
Chapter 1. [no content intended for this page] Model CTF878 Service Manual...
• load updated versions of the meter program into the CTF878 2.1.2 Entering the Service Menu To enter the Service Menu, press the [MENU] key at the lower right of the CTF878 keypad. The Menu Bar replaces the Status Bar at the top of the screen. Press the [ ] arrow key to scroll from the Site Menu to the Service Menu.
• Press [F3] (OK) to confirm that you wish to restore the factory default settings. The program asks for confirmation: “Are you SURE?” Repeat the procedure shown in Step 1 above. The CTF878 replaces the default site file and the global meter settings with the factory default settings, and returns to Operate Mode.
2.1.4 Updating CTF878 Software By using a Windows-based PC with an RS232 connection, CTF878 users can update the meter’s operating software. This software can include the bootloader, FPGA software (timing software for transit-time), instrument software, the meter string file and the Help string file. The updating procedure involves three steps: Setting up the CTF878 in Flash Update mode.
Page 22
(see Figure 11 below). Remove the "CONFIG" jumper from the top two pins. “CONFIG” JUMPER DISPLAY CONNECTOR Figure 11: J13 Header Connector Reinstall the shroud, close the cover, and reapply power to the CTF878. Model CTF878 Service Manual...
Page 23
Figure 12 below). Enter any area code, and then click OK. Figure 12: Location Information Window A New Connection window appears (see Figure 13 below). Enter a name for the New Connection . Click OK. Figure 13: New Connection Window Model CTF878 Service Manual...
Page 24
The Communication Port Properties window appears. Enter the settings shown in Figure 15 below. Click OK. 38400 Figure 15: Communication Port Properties Window The CTF878 is now connected to the PC. In the lower left-hand corner of the Hyperterminal window it reads “Connected” with the connection duration (see Figure 16 below). Model CTF878 Service Manual...
Page 25
• Press [F3] (OK) to confirm that you wish to erase the program. The meter asks for confirmation. Repeat the options shown in Step 1 above. The screens on both the PC and the CTF878 now appear similar to Figure 18 below. Model CTF878 Service Manual...
Page 26
“Send File” entry under “Transfer.” Figure 19: “Ready To Receive Update” Window The “Send File” window appears (see Figure 20 below). Select the replacement software (designated by a .cod extension) from the folder where it has been stored. Model CTF878 Service Manual...
Page 27
Chapter 2. Figure 20: “Send File” Window From the protocol drop-down menu, select “Xmodem.” The software update now begins as the “Xmodem file send” appears (see Figure 21 below). Figure 21: “Xmodem File Send” Window Model CTF878 Service Manual...
Page 28
Chapter 2. After the update is done, the “Final Window” appears as the CTF878 reboots (see Figure 22 below). Figure 22: Final Window 2.1.4.4 Completion When the transfer is complete, to prepare the CTF878 for operation: Disconnect the power to the CTF878.
3.1.1 Introduction For user convenience, the CTF878 offers a Calibration/Test Menu. This menu enables users to perform a variety of functions that might occasionally be required to calibrate the CTF878. Refer to the Service Menu Map, Figure 1-1 on page 11, while moving through the following procedures.
To calibrate an Analog Output, proceed as above, select the desired Calibrate AnalogOut option, and press [ENTER]. The screen appears as shown in Figure 25 on page 21. Proceed as follows: Connect a digital multimeter to the analog output and step through the options using the arrow keys. Model CTF878 Service Manual...
Page 31
You have completed calibrating the outputs. To leave the Output window, • Press [F2] (Cancel) to return to Operate Mode without confirming the changes. • Press [F3] (OK) to confirm the new data. The CTF878 returns to Operate Mode. Model CTF878 Service Manual...
If you pressed [F3] (OK), a clock appears in the middle of the screen with the message “TESTING IO” below it. When the test is complete, the screen returns to Operate Mode. Repeat the process to test additional Analog Inputs. Model CTF878 Service Manual...
Page 33
You have completed calibrating the inputs. To leave the Inputs window, • Press [F2] (Cancel) to return to Operate Mode without confirming the changes. • Press [F3] (OK) to confirm the new data. The CTF878 returns to Operate Mode. Repeat the process to calibrate additional Analog Inputs.
] arrow key to move to the SET box, and then press [ENTER] to set the number of pulses. Compare the value entered with the reading on the frequency counter. They should be very close (within ±1 Hz for frequencies below 1 KHz). Model CTF878 Service Manual...
Chapter 3. You have verified the accuracy of the digital outputs. To leave the Totlizer/Frequency window, press [F3] to exit. The CTF878 returns to Operate Mode. Repeat the process to verify the accuracy of additional Digital Outputs. 3.1.6 Testing and Calibrating RTD’s The Calibration/Test option can be used to test and calibrate RTD’s.
You have completed calibrating the inputs. To leave the Calibrate RTD window, • Press [F2] (Cancel) to return to Operate Mode without confirming the changes. • Press [F3] (OK) to confirm the new data. The CTF878 returns to Operate Mode. Repeat the process to calibrate additional RTD’s.
Page 37
Press [ENTER] or [F3] (OK) repeatedly to check each EEPROM by observing the information for the corresponding option card. When the Test EEPROM information has appeared for every option card present, the screen returns to the Operate Mode. Model CTF878 Service Manual...
Page 38
Chapter 3. [no content intended for this page] Model CTF878 Service Manual...
Page 39
Freq in Hz CCorr Average High High 20 mA Read Read Total Pulses HFHI Elec. System Actual 4 mA HFLO Actual 20 mA 60 Hz LFSP Calibrate Calibrate Calibrate HFGain 50 Hz LFGain Reset Reset Reset Model CTF878 Service Manual...
Introduction The CTF878 is a reliable instrument that is easy to maintain. It will provide accurate flow measurement readings as long as it is operated as described in this manual. If problems do arise with the electronics, the pipe, the fluid, or the transducers, the CTF878 displays an error message specifying the possible problem.
Page 42
Measurement is valid. No error error message Receiver Power CTF878 down. Remove shroud. Make sure Board receiver board is seated correctly and its screws FPGA not loaded, etc. Error are on tight. Power up. Contact factory if problem persists.
Displaying Diagnostic Parameters As part of its measurement menu, the CTF878 offers a list of diagnostic parameters to aid in troubleshooting in the event of flowcell, transducer, or electrical problems. You can select any diagnostic parameter for display as a measurement (see Displaying and Configuring Data, Chapter 2 of the Programming Manual or Chapter 4 of the Startup Guide ).
The gas must be homogeneous, single-phase and relatively clean. Although a low level of entrained particles may have little effect on the operation of the CTF878, excessive amounts of solid or liquid particles will absorb or disperse the ultrasound signals. This interference with the ultrasound transmissions through the gas will cause inaccurate flow rate measurements.
Page 45
(see Programming Site Data, Chapter 1 of the Programming Manual , or Chapter 3 of the Startup Guide ). Contact the factory if you cannot solve a transducer-related problem. Model CTF878 Service Manual...
Page 46
Chapter 4. [no content intended for this page] Model CTF878 Service Manual...
Introduction The electronics console of the CTF878 has been designed to permit easy on-site upgrades and parts replacement. See Figure on page 43 and Figure 1-2 on page 11 for details. The instructions in this chapter, along with a few common tools, are all that is required to perform the following tasks: •...
Chapter 5. 5.1.2 Replacing the Fuse If it has been determined that the fuse in the CTF878 requires replacement, complete the following steps: WARNING! The main power to the CTF878 must be disconnected before proceeding. Open the cover on the electronics console. Remove the two mounting screws and lift the clear plastic shroud out of the electronics console.
Remove the main power to the electronics console. WARNING! The main power to the CTF878 must be disconnected before proceeding. Open the electronics console. Remove the two mounting screws and lift the clear plastic shroud out of the electronics console. Remove the five mounting screws and lift the main aluminum shroud out of the electronics console.
Installing an Option Card The CTF878 flowmeter can accommodate up to six option cards. The option cards are installed into sockets on the rear of the printed circuit board, and they are held in place with a metal bracket. A single metal bracket is used to secure all the installed option cards.
1.8 m (6 ft) of the CTF878. Before taking measurements with the CTF878, refer to Initial Setup , Chapter 3 in the Startup Guide, and to Calibration , Chapter 3 in this Service Manual, for instructions on properly setting up the meter for accurate flow rate measurements.
5.1.6 Replacing the LCD Display The CTF878’s measurements are displayed on a two-pane LCD graphic display panel. The LCD display normally provides years of dependable service, but it is easily field-replaceable when necessary. To replace the LCD display, see Figure on page 43 for the component locations, and complete the following steps: WARNING! The main power to the CTF878 must be disconnected before proceeding.
To purchase the parts listed in Table 2 above or any items not listed in the table, contact the factory for assistance. To make sure the proper components are obtained, be sure to specify the serial number of the CTF878 at the time of purchase.
Page 54
Chapter 5. [no content intended for this page] Model CTF878 Service Manual...
Page 55
INHI INHI +24V +24V INLO INHI +24V STREAM STREAM 100-240 VAC FUSE RS232 1 AMP ANALOG OUT L2/N Backlight Cable Clear Plastic LVD Shroud Fuse Holder LVD Shroud Power Option Connector Mounting Screw, 2 pl Cards Model CTF878 Service Manual...
Page 56
Main Aluminum Shroud w/ 5 mounting screws Clear Plastic LVD Shroud w/ 2 mounting screws Display Shroud Backlight Cable Standoffs, 7pl Display Cable Keypad Cables Conduit Plate Main PC Board Cable Clamps Option Cards Detail C Model CTF878 Service Manual...
Appendix A. Appendix A. Service Record A.1.1 Introduction Whenever any service procedure is performed on the CTF868 flowmeter, the details of the service should be recorded in this appendix. An accurate service history of the meter can prove very helpful in troubleshooting any future problems.
Page 58
Appendix A. Date Description of Service Performed By <Doc Tittle> <Manual Type>...
Page 59
Appendix A. Date Description of Service Performed By <Doc Tittle> <Manual Type>...
Page 60
Appendix A. [no content intended for this page] <Doc Tittle> <Manual Type>...
Page 61
Index Acceleration Limit......... . 3 LCD Display .
Page 62
Index Troubleshooting......... . . 31 Error Messages .
Page 63
Shannon, County Clare Ireland declare under our sole responsibility that the CTF878 Ultrasonic Flowmeter to which this declaration relates, are in conformity with the following standards: • EN 61326:1998, Class A, Annex A, Continuous Unmonitored Operation • EN 61010-1:1993 + A2:1995, Overvoltage Category II, Pollution Degree 2 following the provisions of the 2004/108/EC EMC Directive and the 2006/95/EC Low Voltage Directive.
Page 64
Shannon, County Clare Ireland déclarons sous notre propre responsabilité que les CTF878 Ultrasonic Flowmeter rélatif á cette déclaration, sont en conformité avec les documents suivants: • EN 61326:1998, Class A, Annex A, Continuous Unmonitored Operation • EN 61010-1:1993 + A2:1995, Overvoltage Category II, Pollution Degree 2 suivant les régles de la Directive de Compatibilité...
Page 65
Shannon Industrial Estate Shannon, County Clare Ireland erklären, in alleiniger Verantwortung, daß die Produkte CTF878 Ultrasonic Flowmeter folgende Normen erfüllen: • EN 61326:1998, Class A, Annex A, Continuous Unmonitored Operation • EN 61010-1:1993 + A2:1995, Overvoltage Category II, Pollution Degree 2 gemäß...
Page 67
Customer Support Centers U.S.A. The Boston Center 1100 Technology Park Drive Billerica, MA 01821 U.S.A. Tel: 800 833 9438 (toll-free) 978 437 1000 E-mail: mstechsupport@bakerhughes.com Ireland Sensing House Shannon Free Zone East Shannon, County Clare Ireland Tel: +353 61 470200 E-mail: mstechsupport@bakerhughes.com Copyright 2021 Baker Hughes company.
Need help?
Do you have a question about the CTF878 and is the answer not in the manual?
Questions and answers