forward to one position. When the figure displayed is same as the capacity of
the standard weight, press the [On/Off] key.
For example: If a 10 kg weight is used instead of full capacity:
After you process till the step (9) as above, press the [On/Off] key.
The indicator will show
automatically jumped from
[Hold] key three times. Now, the
and the display will be changed to
press the [On/Off] key immediately.
Next the display will show the current weight on the pan.
10. The Calibration is now complete.
11. Set the span jumper on the PCB to OFF position and re-seal the indicator.
12. If the selections made during calibration are not as described as above, the
scale may still work but will not show accurate results. In such cases, re-do
the calibration as shown above.
© Adam Equipment Company 2006