Usage instructions Adapter & Adaptation
Then insert the enclosed cable ties into the
internal grooves on the centring semi-shells
designed for this purpose. When tightening the
cable ties, ensure that their closures are at the
back so that they do not affect the adapter
when it is being adapted. The cable ties ensure
the proper position of the clamping bracket
fixing hooks and thus prevent flapping.
87 mm
Cable ties on centring
semi-shells
Figure 24: Fully assembled centring semi-shells
Now assemble the centring column with the
terminal clamp on the back frame cross tube.
Before you tighten the M6 fixing screws (Metric
size 5 mm), you must adapt the adapter as
described in Chapter 5.3 and turn the centring
column by twisting the terminal clamp on the
back frame cross tube to the correct angle.
With the adapted adapter, you must then
tighten the fixing screws with a tightening
torque of 7 Nm. Ensure that the centring col-
umn remains located at the centre of the back
frame cross tube.
Centring column
Figure 25: Centring column with terminal clamp
5.2

Assembly work on adapter

The adapter has adaptation elements that are
pushed onto the fixing elements available or
assembled on the wheelchair and locked in the
adaptation position:
Figure 26: Adaptation elements on the adapter
The adjustment option on the clamping bracket
of 105 mm in seven 15 mm steps as well as
the infinite horizontal adjustment of the cen-
tring column mount enables the horizontal
position of the adapter to be set on the wheel-
chair and adjusts the adapter to the geometry
of the wheelchair. Adapt the adapter to the
wheelchair as described in Chapter 5.3 and
check the necessary adjustment work.
To position the clamping bracket horizon-
tally, the M8 fixing screw (Metric size 5 mm)
must be released and tightened again in the
right thread hole in the fixing rail at 17 Nm and
protected with locking fluid.
Clamp
Figure 27: M8 fixing screw to horizontally position
the clamping bracket
At the end of the adapter is the centring
column mount; its position is infinitely ad-
justable.
In the centre of the adapter is the clamping
bracket that can be positioned in a range
of 105 mm in seven 15 mm steps and the
gap between it and the adapter tube can
be adjusted using spacers.
Clamping bracket with clamp
levers and fixing hooks
Pull loop on
Centring column
clamp levers
mount
Fixing hooks
M8 fixing screw
Spacers
10