Standard Report For Iso/Iec 16022 "Standard / Id-Genius" Decoder - Siemens SIMATIC MV500 Operating Instructions Manual

Hide thumbs Also See for SIMATIC MV500:
Table of Contents

Advertisement

Image processing
3.2 Code verification
Note
Verification error
Whenever a verification error occurs, all quality grades are output as "E" and the error number
is output instead of the quality values. For information on the meaning of error numbers, see
the section "Read and verification error messages (Page 229)".
3.2.9.1

Standard report for ISO/IEC 16022 "Standard / ID-Genius" decoder

Table 3- 10
Field name (String input)
Verification type (Qt)
Evaluation (Qs)
Overall grade (Q0)
Contrast grade (Q1)
Contrast (q1)
Axial non-uniformity grade (Q2)
Axial non-uniformity (q2)
Print growth grade (Q9)
Print growth X (q9)
Print growth Y (q9)
Unused error correction grade (Q4)
Unused error correction (q4)
Cell size (q10)
ECCLevel (q14)
Height (q15)
Width (q16)
Angle (q17)
68
Output parameters: ISO/IEC 16022 "Standard" decoder
Value
4
If the exposure settings do not conform to the
standard or uncalibrated verification is active: E
3 (Good), 2 (Fair), 1 (Poor)
4 ... 0
4 ... 0
0 ... 100
4 ... 0
0.00 ... 1.00
4 ... 0
-100 ... 100
-100 ... 100
4 ... 0
0.00 ... 1.00
0 to 99.9 mil
Uncalibrated verification: Output in pixels [px]
200
0 to 9.99 inches
Uncalibrated verification: Output in pixels [px]
0 to 9.99 inches
Uncalibrated verification: Output in pixels [px]
0 ... 360°