Auditor tab
This tab provides the export of full data as Auditor readable smr format.
Figure 2.13: Export - Auditor
Additional Information tab
This tab allows to put additional information showing up in all export files. Especially a measurement time
can be applied manually using the
Figure 2.14: Export - Additional Information
2.4 Calibrate
A list of the approved calibrators and the accuracy classes achieved is given in section 4.2. Follow the
instructions below to calibrate the device at 1 kHz.
2.4.1 Calibrate using Tango-Utilities
1. Connect Tango_Plus to the PC.
2. Start Tango-Utilities.
3. Insert the microphone into the calibrator and activate the calibration signal. The reference calibration
level is 94 dB for this methode.
4. Press the
button in Tango-Utilities. "Calibration activated" appears in the status bar. To cancel
CAL
the calibration procedure, press the stop button.