Introduction - Siemens SITRANS F M Verificator Operating Instructions Manual

Table of Contents

Advertisement

Introduction

The SITRANS F M MAGFLO Verificator consists of:
● a stand alone Verificator to measure a number of selected parameters in the flow sensor
and a transmitter which affects the integrity of the flow measurement.
● a Windows based PC programme enabling printing and management of verification reports.
20 verifications can be stored in the Verificator before uploading to a PC or printing of a report
if necessary.
Verification of a SITRANS F M MAGFLO flowmeter consists of the following test routines:
● insulation test of the entire flowmeter system and cables
● test of sensor magnetic properties
● transmitter gain, linearity and zero point test
● digital output test
● analog output test
When connected to a PC, a full verification report containing all test results can be printed.
The Verificator can be used with SITRANS F M MAGFLO
1100, MAG 1100 FOOD, MAG 3100, MAG 3100 W and MAG 5100 W used together with
transmitters MAG 5000 or MAG 6000.
NOTICE
Short circuit
A short circuit to earth will damage the Verificator.
NOTICE
Use in a domestic environment
This Class A Group 1 equipment is intended for use in industrial areas.
In a domestic environment this device may cause radio interference.
SITRANS F M Verificator
Operating Instructions, 07/2019, A5E45134218-AB
flowmeters with sensor types MAG
®
1
5

Advertisement

Table of Contents
loading

Table of Contents