Re-Calibration - Siemens SITRANS F M Verificator Operating Instructions Manual

Table of Contents

Advertisement

Important
Please check that an earth wire has been mounted between connection plate SHIELD and
connection box. If it is missing, please mount an earth wire of at least 1 mm
7.2

Re-calibration

Assuring the quality of the MAG Verificator measurements is crucial. In order to achieve the
best measurements and thereby the correct decision based upon it, it is required that the MAG
Verificator is send in for re-calibration once every year. This both establishes (metrological)
traceability for the measurements and fulfills the requirements of ISO 9001.
Also be aware, that the re-calibration date appears on the MAG Verificator display and the
printed or stored certificate.
Re-calibration of the MAG Verificator is offered at our Instrumentation Flow Factory in France.
Address is:
Siemens SAS
PD PA MF-H CSC
Customer Care Center
1 Chemin de la Sandlach
67506 Haguenau Cedex
France
Following code numbers can be used:
Description
STANDARD
package
ADVANCED
suitcase package
ADVANCED
cable package
ALL-IN-ONE
package
Note
Please use the VSR procedure for sending in equipment to the factory.
SITRANS F M Verificator
Operating Instructions, 07/2019, A5E45134218-AB
WARNING
Scope of service
● Device check
● Calibration
● Function test
● Check and if broken exchange connector and relay
● STANDARD package
● Exchange of suitcase and foam inserts
● STANDARD package
● Exchange of adapter cable incl. connector
● STANDARD package
● ADVANCED Suitcase and cable package
Diagnostics and troubleshooting
7.2 Re-calibration
.
2
MLFB
9LA1110-8QT10-0BA0
9LA1110-8QT10-1BA0
9LA1110-8QT10-2BA0
9LA1110-8QT10-9BA0
35

Advertisement

Table of Contents
loading

Table of Contents