Quick start guide CellaTemp PA
Electrical connection
The pyrometer is supplied with low voltage 24V DC (18 ... 32V DC).
The pyrometer may only be installed by a skilled, qualified
electrician. Do not connect the instrument while the voltage
supply source is turned on. Please observe international
safety regulations at all times
► Switch to neutral and verify absence of voltage
► Connect the instrument according to the following schematic.
PIN 1
WH (white)
PIN 2
BN (brown)
PIN 3
GN (green)
PIN 4
YE (yellow)
PIN 5
GY (grey)
PIN 6
PK (pink)
PIN 7
BU (blue)
PIN 8
RD (red)
The pyrometer must be protected against high voltage and strong
electromagnetic fields. Use a shielded cable. The shield must be
connected with the connector housing.
Use a flyback diode when switching inductive loads.

Installation

Install the pyrometer in a location where it will not be unnecessarily ex-
posed to smoke, ambient heat or water vapour.
1. Select a suitable location for the pyrometer. The following criteria
must be observed:
L+ (Power supply 24V DC)
Digital in- / out 1
Analogue output 2/ analogue input
Analogue output 1
Serial interface RS 485 (A) /Tx-/D-)
Serial interface RS 485 (B) /Tx+/D+)
Digital in- / output 2
L- (GND)