REPLACING THE BATTERY DOOR

If the battery door
is removed or
needs replacing,
simply snap the
door onto the bar
that is located on
the pump.

POWER UP

Press and hold the power switch to
turn the pump on. The pump starts
the power up sequence during which it
performs various self-tests and tests
for alarm conditions.
If any issues are found while the
pump is performing the self tests,
alarms will sound
If you believe there is a problem,
remove the pump from service
and contact Smiths Medical
Customer Service
CADD
-Solis VIP Pump
®
AA batteries
with AC adapter
Rechargeable
battery with
AC adapter
CAUTION: If the power up results
in an error message indicating that
the protocol library was lost, do
not proceed with using the pump.
Follow your facility's procedures for
downloading protocol libraries.

POWER DOWN

Press the power switch to turn the
pump off. You must acknowledge the
"Power down?" prompt by selecting
Yes. The pump then powers down.

CASSETTES

The cassette is part of the CADD™
medication cassette reservoir or
CADD
administration set that attaches
®
to the bottom of the pump.
Notes:
A CADD
administration or
®
extension set with free-flow
protection must be used
A CADD
high volume
®
administration set is required
for rates above 250 mL/hr
p. 27