Note: In step and taper therapies,
you may not set a delayed start after
an infusion has been started.
To Set a Delayed Start:
1.
From the Tasks menu, press ▲
or ▼ until Set Delayed Start is
highlighted, then press Select
2.
Unlock the keypad, if required
3.
Press ▲ or ▼ until the desired start
time appears on the screen and
select Confirm
4.
Confirm the new start time by
selecting Yes
Next Dose Start Time
Note: This task is available for
intermittent therapies only.
A Next Dose Start Time may be
programmed to delay the start of the
next dose of an infusion. However if a
dose is in progress, this cancels the
remainder of the current dose. To avoid
interrupting the dose in progress,
adjust the next dose start time when
the pump is in KVO.
Note: During dosing, the home screen
shows "Dosing". If a dose is stopped
in progress, the home screen shows
"Interrupted." If the infusion is in KVO,
"Between Doses" appears on the
home screen.
The Next Dose Start Time may also be
set to delay delivery of the first dose
of the infusion by up to 96 hours. The
pump must be running on the selected
date and time in order for delivery
to begin.
p. 14
To Set the Next Dose Start Time:
1.
Stop the pump if it is running. If a
dose was delivering when the pump
was stopped, the "Interrupted"
screen appears
Note: If a dose is in progress, this
cancels the remainder of the dose.
2.
From the Tasks menu, press ▲ or
▼ until Set Next Dose Start Time is
highlighted, then press Select
3.
Unlock the keypad, if required
4.
Press ▲ or ▼ until you reach the
time that you want the next dose
to begin. The length of the delay is
listed below the time.
Select Confirm
5.
Confirm the new start time and
length of the delay. Select Yes

EDIT / VIEW DELIVERY SETTINGS

The delivery settings are patient-
specific parameters of a therapy that
are directly related to the drug being
infused and can be edited within limits
established in the protocol. The CADD
Solis system administrator establishes
the initial values of the parameters, any
applicable programming units and drug
concentration, other parameter limits,
and which parameters can be viewed
and/or edited. A security code (or the
pump key, if enabled) are required to
edit the parameters.
Note: Editing individual delivery
settings in Step or Taper mode
will reset the infusion back to
the beginning.
-
®