Smiths Medical CADD Solis VIP 2120 Clinicians Manual page 20

Ambulatory infusion pump
Table of Contents

Advertisement

Dose Rate
This setting is for review only and
shows the rate at which the dose is
delivered based on the programmed
dose volume and dose duration.
The maximum allowable rate is
500 mL/hr. Rates above 250 mL/
hr require a CADD
administration set
Next Dose
The time the next dose is scheduled
to begin. The pump must be
running on the selected date
and time in order for the dose to
begin. The next dose time can be
programmed at any time to delay
the start of the next dose, but if a
dose is in progress, this cancels the
remainder of the dose
KVO Rate
The KVO or "keep vein open" rate
is optional. It allows the delivery of
a minimal amount of drug to help
maintain catheter patency. The KVO
rate is active between doses or until
a dose begins if the next dose start
time has been set to a value other
than "immediate"
Reservoir Volume
The volume of fluid contained in
the reservoir. The reservoir volume
value decreases as the pump is
primed or delivers fluid
p. 20
high volume
®
Stopping the Pump
During an Infusion
Stopping the pump between doses
does not affect the start time of
subsequent doses. Stopping the
pump while a dose is in progress
shifts all subsequent doses by the
amount of time the pump is stopped
Resuming the Dose
To resume delivery of a dose
stopped in progress, simply
restart the pump by pressing
stop/start
Resetting the Cycle
To reset the cycle, stop the pump
between doses, and reprogram
the next dose start time. When a
dose is stopped in progress, the
"Interrupted" screen appears. To
cancel the current dose and reset
the cycle, reprogram the next dose
start time through the Tasks menu.
The next cycle will start at the time
you select
Clinician Guide