Summary of Contents for Smiths Medical Graseby 500
Page 1
Smiths Medical cannot be held responsible for the accuracy and any resulting incident arising from information that has been extracted from this publication and compiled into the users documentation.
Page 2
® Graseby Model 500 ® and Graseby Micro 505 Volumetric Infusion Pump Technical User Manual...
Page 3
Company number 362847 Trademarks and acknowledgements: Graseby is a trademark of the Smiths Medical family of companies. The symbol ® indicates that it is registered in the U.S. Patent and Trademark office and certain other countries. The products described are covered by one or more of the following Patent Nos.: U.S.
Page 4
The pump must be referred to a suitably qualified WARNING: When delivering drugs in the technician or returned to Smiths Medical in epidural space, use only those medications order to have the fault rectified. specifically indicated for epidural use.
Page 5
20. WARNING: Check the Secondary set carefully, 28. WARNING: While Smiths Medical have taken since an occlusion above the pump on the all reasonable steps to ensure that the pump Secondary line could cause the Primary fluid...
Page 6
W a r n i n g s a n d C a u t i o n s S m i t h s M e d i c a l Cautions Cautions tell you about dangerous conditions that can occur and cause damage to the pump if you do not obey all of the instructions in this manual.
Page 7
S m i t h s M e d i c a l W a r n i n g s a n d C a u t i o n s V o l u m e t r i c I n f u s i o n P u m p s...
S m i t h s M e d i c a l C o n t e n t s C o n t e n t s Who should read this manual ................. 1 Introduction ....................... 1 Contents of the box ................... 2 Checking labels and serial number ..............
Page 9
C o n t e n t s S m i t h s M e d i c a l V o l u m e t r i c T e c h n i c a l U s e r G u i d e...
Manual, available on request from information are not needed by most Smiths Medical. users, so they are separated into this stand-alone manual. W h o s h o u l d r e a d t h i s m a n u a l You should be familiar with the use of •...
Epidural Label Set for the Volumetric replacement parts. Infusion Pump (part number TPF-00306) is available from Smiths Medical. V o l u m e t r i c I n f u s i o n P u m p...
T e c h n i c a l U s e r M a n u a l S m i t h s M e d i c a l F u n c t i o n a l i n s p e c t i o n s a n d t e s t s Perform the following tasks when you Power up and self test have taken the pump out of the...
Page 14
If any of the following faults occur prime a ‘Standard Adult Set, 8C-820’ with a new pump, contact your (non-checkvalve 20 drop/mL) Smiths Medical Service Department. administration set Product Code • if any LEDs do not light in steps 1, 3 21-0346-25 for the following tests.
Page 15
If you have a new pump and it does 6. Press the Run key and then press not behave as described here, the Options key. Verify that the contact your Smiths Medical message display shows the time Service Department. remaining with a value like the...
Page 16
If you have a new pump and it does and push it into its slot. Close the not behave as described here, clamp above the cassette. Then close contact your Smiths Medical the door and silence the alarm when Service Department. this message appears:...
T e c h n i c a l U s e r M a n u a l S m i t h s M e d i c a l S e t t i n g u p Battery charging Enabling limits Before using the pump for clinical use,...
S m i t h s M e d i c a l T e c h n i c a l U s e r M a n u a l R e g u l a r t e s t i n g In addition to the regular testing requirements listed below, Smiths Medical recommends that a qualified...
T e c h n i c a l U s e r M a n u a l S m i t h s M e d i c a l M e n u s There are several menus on the pump, The Biomedical menu some of which have an overlap of This menu is the one that also exists in...
S m i t h s M e d i c a l T e c h n i c a l U s e r M a n u a l Te c h n i c i a n m e n u This menu allows you to define how the You cannot move back through the pump behaves by setting values or by...
T e c h n i c a l U s e r M a n u a l S m i t h s M e d i c a l Technician parameters The numbers on the left of this table show the number of Options key presses to reach the parameter.
S m i t h s M e d i c a l T e c h n i c a l U s e r M a n u a l Using Technician parameters Most parameters enable or disable can be infused for a Primary infusion.
Page 23
T e c h n i c a l U s e r M a n u a l S m i t h s M e d i c a l Enabling this feature allows users to Secondary stop use the Options key to put the pump With this parameter enabled, the pump into a mode in which they can change...
Page 24
S m i t h s M e d i c a l T e c h n i c a l U s e r M a n u a l When the pump is in use, if the date set Backlight is reached or exceeded, then the This parameter sets the default for how...
T e c h n i c a l U s e r M a n u a l S m i t h s M e d i c a l are displayed for security protection. If B i o m e d i c a l m e n u you make a mistake, you must switch the pump off to enter the correct ID.
S m i t h s M e d i c a l T e c h n i c a l U s e r M a n u a l B i o m e d i c a l m e n u p a r a m e t e r s The numbers in the left hand column of this table indicate the number of Options key presses to reach the parameter.
Page 27
T e c h n i c a l U s e r M a n u a l S m i t h s M e d i c a l Screen Notes Shows the current rate Rates of 300, 600, setting of the serial port.
S m i t h s M e d i c a l T e c h n i c a l U s e r M a n u a l S e r v i c e F u n c t i o n s m e n u The Service Functions menu allows you Using the Service Functions menu to set the date and time on the pump...
S e r v i c e F u n c t i o n s m e n u p a r a m e t e r s Screen Notes Information for Smiths Medical Service Department use. Use this to set the time 1.
S p e c i f i c a t i o n S m i t h s M e d i c a l S p e c i f i c a t i o n s General Weight 5 kg (11 pounds).
Page 33
Test solution Sterile water or normal saline at room temperature (70°F ±5°/21°C ±3°). Smiths Medical standard (primary), 20 drops/mL, non-checkvalve administration set (8C820). Testing conditions Model 500 Fluid level in the solution container 46 cm (18 inches) above top of the pump, rate set at 999 mL/h and volume to be infused of 49 mL.
Page 34
S p e c i f i c a t i o n S m i t h s M e d i c a l Power AC power supply Internally configured for either 100-120 V AC, 200 mA, 50/60 Hz. 220-240 V AC, 80 mA, 50/60 Hz.
Page 35
S p e c i f i c a t i o n S m i t h s M e d i c a l Primary and Secondary Infusions: Rate range Model 500 Range Increment 0.1 to 99.9 mL/h 0.1 mL/h 1 to 999 mL/h 1 mL/h...
Page 36
S p e c i f i c a t i o n S m i t h s M e d i c a l Volume Over Time Infusions: Rate range Model 500 Range Increment 0.1 to 99.9 mL/h 0.1 mL/h 1 to 999 mL/h 1 mL/h...
Page 37
S p e c i f i c a t i o n S m i t h s M e d i c a l Drug amount modes/range Mode Range Increment MG drug/bag 0.01 to 99.99 0.01 0.1 to 999.9 1 to 99999 Gm drug/bag 0.01 to 99.99...
Page 38
0.1 to 3.0 mL/h Accessories For a complete list of Administration Sets, please contact Smiths Medical or your local distributor. V o l u m e t r i c I n f u s i o n P u m p...
Page 39
S p e c i f i c a t i o n S m i t h s M e d i c a l Symbols used on the pump Front panel symbols Inside battery door symbols Battery is charging/mains Attention: dangerous power applied.
European Council Directive 93/42/EEC concerning medical devices. 0473 The number 0473 identifies the Notified Body under which the Quality Systems operated within Smiths Medical are assessed. Disposal When the time comes to dispose of the pump, its batteries, or any of its accessories, do so in the best way to minimise any negative impact on the environment.
S p e c i f i c a t i o n S m i t h s M e d i c a l Tr u m p e t c u r v e s The curves were developed while The trumpet curve represents the worst testing the Model 500 using a Smiths case rate error in any given observation...
Page 42
S p e c i f i c a t i o n S m i t h s M e d i c a l V o l u m e t r i c I n f u s i o n P u m p S —...
Page 43
S p e c i f i c a t i o n S m i t h s M e d i c a l S — 1 2 V o l u m e t r i c I n f u s i o n P u m p...
Page 44
S m i t h s M e d i c a l T e c h n i c a l U s e r M a n u a l - I n d e x I n d e x AC cord retainer 4 Damaged pump 2 Acceptance test 4...
Page 45
I n d e x - T e c h n i c a l U s e r M a n u a l S m i t h s M e d i c a l Occlusion Alarm default rate S - 7 approx values of settings S - 7 rate entry parameter 13...
Page 46
S m i t h s M e d i c a l T e c h n i c a l U s e r M a n u a l - I n d e x Service Functions menu description 19 Unpacking pump 2 overview 10...