Giving a Clinician Bolus
A clinician bolus may be delivered only
when the pump is running. It allows
delivery of a specific amount of drug,
such as a loading dose. A clinician
bolus cannot be started while a PCA
dose is in progress. The clinician bolus
increases the given amount, but does
not add to the dose counters or to the
delivery limit. A clinician bolus may be
stopped in progress.
Note: The maximum clinician bolus
may be limited by the settings in the
protocol, which is determined by the
CADD
-Solis system administrator.
®
To start a clinician bolus:
1.
Make sure the pump is running
2.
From the View Advanced
Tasks menu, press ▲ or ▼
to highlight Give Clinician
Bolus and press Select
3.
Unlock the keypad
4.
Press ▲ or ▼ to change the value of
the clinician bolus. Select Deliver
5.
If you enter a value outside the soft
limit, a screen appears asking you
to confirm the soft limit override.
Select Confirm to deliver the bolus
6.
The screen shows the decreasing
amount as the bolus is delivered.
You may stop the bolus at any time
by selecting Stop Bolus
7.
When the bolus has been delivered,
the screen says, "Clinician bolus
complete." Select OK
CADD
-Solis VIP Pump
®
Start a PCA Dose
If a PCA dose is programmed, the
patient may start a PCA dose while the
pump is running. The amount delivered
is added to the amount provided by
the continuous rate. Each time the
patient requests a PCA dose, the pump
automatically adds it to the Given and
PCA Dose Counters report.
The remote dose cord can be attached
to the pump and provided to the patient
as an alternative to pushing the PCA
dose key when requesting a PCA dose.
The LED on the remote dose cord
indicates PCA dose status:
Off: A PCA dose is not available
Flashing: A PCA dose is available
On: A PCA dose has been requested
and delivery has started
Notes:
A PCA dose cannot be started while
another PCA dose or a clinician
bolus is in progress
The PCA dose key on the pump is
inactive when the remote dose cord
is connected to the pump
If the display has automatically
blanked, pressing the PCA dose
key turns the display back on and
delivers a PCA dose, if available