To Start a PCA Dose:
Press the PCA dose key on the pump
or the remote dose cord. If a dose is
available, it will start infusing.
Stopping a Clinician Bolus or PCA Dose
To Stop a Clinician Bolus:
While the clinician bolus is delivering,
the Clinician Bolus screen appears on
the pump and shows the decreasing
amount of the bolus remaining to be
delivered. You may stop the clinician
bolus from the screen by selecting
Stop Bolus.
To Stop a PCA Dose:
From the "PCA dose started" screen
you may stop the PCA dose while it is
delivering by selecting Stop Dose.
Or
If the pump is on the home screen,
press the Stop/Start button.
You will first be asked if you want to
stop the dose, select Yes. Then you will
be asked if you want to stop the pump,
select No.

CONTINUOUS DELIVERY MODE

Continuous delivery allows the
infusion of drug fluid at a constant,
programmed rate.
Home Screen
The delivery mode and status of the
infusion are indicated on the home
screen with messages appearing in
text, a graphic in a shape unique to
the therapy, and with the colors green,
red, and grey.
p. 18
Examples specific to the continuous
delivery mode are:
The pump is running with a
programmed delayed start
The pump is stopped. It will
deliver at a continuous rate
once it is started
Programming Screens
Continuous Rate
The desired continuous rate of
medication delivery. Rates above
250 mL/hr require a CADD
volume administration set
KVO Rate
The KVO or "keep vein open" rate
is optional. It allows the delivery
of a minimal amount of drug to
help maintain catheter patency. If
a KVO rate is programmed, it will
only be active if a delayed start
is programmed. The pump will
infuse at the KVO rate until the
programmed infusion begins
Reservoir Volume
The volume of fluid contained in
the reservoir. The reservoir volume
value decreases as the pump is
primed or delivers fluid
Start Time
The time the next infusion will
begin. It is displayed only if the
delayed start is programmed.
The pump must be running on
the selected date and time in order
for the infusion to begin
high
®