X-Ray Tube Technical Information; X-Ray Generator; Imaging System Technical Information; Image Receptor - Hologic Trident HD User Manual

Table of Contents

Advertisement

Trident HD Specimen Radiography System User Guide
Appendix A: System Specifications
A.5

X-ray Tube Technical Information

Source-to-Image Distance (SID)
Focal Spot
Tube Voltage
Duty Cycle
Anode Material
X-Ray Window
A.5.1

X-Ray Generator

kV Range
mA
mAs
A.6

Imaging System Technical Information

A.6.1

Image Receptor

Fluid Ingress
Active Imaging Area
Dynamic Range and Linearity
Uniformity
Image Pixel Size
Digital Image Receptor MTF
Page 76
43 cm ±0.5 cm (16.9 inches ±0.2 inches)
no more than 55 microns
20 kV to 50 kV
16%; 20 second exposure every 125 seconds
Tungsten
Beryllium 200 μm
20 kV to 50 kV
1 mA
20 mAs maximum
The Image Receptor is protected from accidental spillage on the Cabinet
floor.
Mid-field Detector: 12 cm x 14 cm nominal
Full-field Detector: 20 cm x 20 cm nominal
Detector Subsystem response is linear with linearity of 0.999 over a
dynamic range of 400:1 in x-ray exposure.
Detector Subsystem can correct pixel-to-pixel gain variations.
70 μm
Modulation Transfer Function (MTF) not less than 40%
at 7.1 lp/mm
MAN-05478-002 Revision 003