1760 Users
Change #3,
On page 6-4, in the first table, replace footnotes 2 and 3 with:
2
sensor + basic unit 0.1 % of Udin = 230 V P-N in accordance with IEC61000-4-30 Class A
- for values <10 % and >150 % of Udin 0.15 % of rdg ±0.48 V
3
sensor + basic unit 0.1 % of Udin = 480 V and 600 V P-N in accordance with IEC61000-4-30 Class A
- for values <10 % and >150 % of Udin 0.15 % of rdg ±0.8 V
2/18
558
5
Manual Supplement