7.7 Starting up procedure
® ® ® ® ®
with SENSORPROM
flow memory unit
7.8 Starting up procedure
without SENSOR-
® ® ® ® ®
PROM
flow memory
unit
Wet calibration
Theoretical calibration
7. Starting up
1. Switch on the signal converter SONO 3000. The meter will automatically run through a self-test
routine. During the self-test the display will show the text ROM TEST, RAM TEST and INITIA-
LIZING. The self-test is completed when the display starts showing the volume flow rate.
2. If the error symbol is displayed, go to section "Trouble shooting".
3. Refer to the CONVERTER SETUP MODE, OUTPUT SETUP to set the required output signals.
1. Switch on the signal converter SONO 3000. The meter will automatically run through a self-test
routine. During the self-test the display will show the text ROM TEST, RAM TEST and
INITIALIZING. The self-test is completed when the display starts showing the volume flow rate.
2. When the signal converter detects no SENSORPROM
mm sensor settings. Error 42 is shown for a short while. The signal converter will display the
corresponding flow rate.
3. For correct flow measurement the following parameters must be set:
CONVERTER SETUP MODE:
BASIC SETTINGS:
The dimension-dependent settings are listed in table 2 (previous page).
OUTPUT SETUP:
Set the required output signals.
SENSOR CHARACTERISTICS:
CALIBRATION CHOICE
Choice: WET, AUTO
If WET:
CAL.FACTOR
APPLICATION PARAMETERS
Cable length
PIPE GEOMETRY
PIPE DIAMETER
NO. OF TRACKS
TRACK LENGTH
If AUTO:
All PIPE GEOMETRY data and APPLICATION PARAMETERS are required
UNIT
PIPE DIAMETER (Internal)
ROUGHNESS (Steel pipes 0.0004 m typically)
NO. OF TRACKS
TRACK ON/OFF
TRACK LENGTH
TRACK ANGLE
TRACK DISPLACEMENT
VISCOSITY UNIT
VISCOSITY
4. To set required measuring ranges go to BASIC SETTINGS.
5. If the error symbol is displayed, go to section "Trouble shooting".
6. Zero adjustment is only necessary if the user requires a high accuracy at very low flows.
7. To set the required output signals go to the CONVERTER SETUP and OUTPUT SETUP
MODES.
The procedure described above is only required when setting up the signal converter
for the first time.
The signal converter will retain the settings even in case of power failure.
SFIDK.PS.029.Q3.02
®
unit, it will configure itself to a DN 400
49