CLEANING THE CABINET FILTER
1. Shut off the cabinet Power Supply
2. If equipped with a side component cabinet, use the provided 8mm Allen wrench to remove the mechanical compartment
front panel. For cabinets with components mounted in the top, remove the access or top panel (depending on model)
3. Turn off the switch for HEPA Fan and unplug the power cord
4. Slide the HEPA housing completely out of the cabinet and remove the lid by undoing two latches and one screw
5. Twist the Filter Assembly counterclockwise to unlock the old HEPA Filter
6. If replacing the HEPA Filter along with its Pre- and Post-Filters, go to STEP 7
a. Remove the Pre-Filter Sleeve by sliding it off the HEPA Filter and replace with the new one
b. Remove the Post-Filter from the opening at the bottom of the HEPA Filter. To insert the new Filter, roll it up,
and slide it back into the HEPA Filter
c. Replace the O-Ring on the HEPA Filter
d. Place the old Filters (to be discarded) in a bag and remove to a non-biohazard area
7. Insert the new HEPA Assembly into the housing, and with downward pressure, twist the Filter Assembly clockwise to
relock the Seal and Filter into place
8. Replace the lid to the HEPA Housing and slide it into the cabinet
9. Plug in Fan and turn on the Power Switch to ON
10. If using EndoTrack, go to the Service Screen to reset the HEPA replacement timer. Refer to the EndoTrack User Guide
11. Replace the mechanical compartment front panel, top panel, or shroud. Store the Allen wrench in the hold on the top
side of the cabinet
12. Turn on cabinet Power Supply
Cabin HEPA Filter Box With Top Removed
Twist Cap (To Left) To Access HEPA Filter
8
|
ENDOSCOPE DRYING CABINETS HEPA FILTER
Cabin HEPA Filter
USER GUIDE
VOC Carbon Filter
(800) 665-3760