ELECTRICAL SUPPLY
• North America
120 Vac, 60 Hz, 15 Amp Standard Deluxe Receptacle
• Typical Full Load Amperage
~1 Amps
• It is recommended to plug any cabinet into a dedicated circuit
• Units with EndoTrack should be plugged into an outlet connected to the emergency back-up circuit
• Disconnect power whenever servicing the cabinet, by unplugging the power cord
ENVIRONMENTAL CONDITIONS
The cabinets have been designed for use within medical facilities. They are designed to be safe within the
following conditions:
• Indoor Use ONLY
• Temperature Range: 15-35°C
• Maximum Relative Humidity: 80% up to 25°C and decreasing linearly to 50% relative humidity at 35°C
CLEANING OF CHANNEL PURGE TUBING SETS
• At the hospital's discretion, the channel purge tubing set and connectors, can be cleaned by Steam Sterilization,
hospital approved Enzymatic detergents or High Level Disinfection
• Steam Sterilization: Tubing sets are compatible with steam sterilization at the following parameters:
270°F ( 32°C), four-minute cycle. Tubing sets should be allowed to dry for 30 minutes
connecting to an Endoscope
• The tubing sets should be washed using a hospital approved Enzymatic cleaner prior to the first use. This is done
by flushing, soaking and rinsing, following the Enzymatic manufacturer's IFU. After the tubing sets have been
properly rinsed, they can be connected to the quick connect manifold in the cabinet to allow the interior of the
tubing to be air purged dry for 20 minutes prior to connecting to an Endoscope
• Approved High Level Disinfectants include Peracitic Acid, Hydrogen Peroxide, Gluteraldehyde, and OPA base
disinfectants. Follow the respective HLD manufacturer's IFU for proper use. After the tubing sets have been
properly rinsed, they can be connected to the quick connect manifold in the cabinet to allow the interior of the
tubing to be air purged dry for 20 minutes prior to connecting to an Endoscope
• Tubing sets and connectors should be checked regularly for cracks and loss of tubing flexibility. If any such flaws
are found, they should be replaced immediately
Based on independent third-party testing.
*
www.logiquip.com
ENDOSCOPE DRYING CABINETS HEPA FILTER
USER GUIDE
prior to
*
|
3