Page 1
USE & INSTALLATION GUIDE Surgical Scrub Sinks USE & INSTALLATION GUIDE Read and understand all of the instructions and safety information CAUTION in this document before assembling and installing this product. (800) 665-3760 www.logiquip.com customersupport@logiquip.net...
• SS64 – 2-Station Scrub Sink • SS96 – 3-Station Scrub Sink PURPOSE OF THIS GUIDE This user guide details the installation, operation maintenance and cleaning of LogiQuip Surgical Scrub Sinks It also contains general specifications, warnings and cautions PLEASE KEEP THIS GUIDE...
Severe shock hazards shall be a lightning bolt in a red triangle. which might result from the combination of this product with other products, whether supplied by LogiQuip or by or by other manufacturers, unless such a combination has been specifically endorsed by LogiQuip...
SCRUB SINKS USE & INSTALLATION GUIDE IN-LINE FLOW SWITCH TIMER CONTROLLER (S0130-01) – SPECIFICATIONS, SWITCH RATING AND WIRING DIAGRAM NOTE: This feature (Fig 1) is available only on Infrared Activated Scrub Sinks Service Compatible liquids Housing: brass; Piston: polysulfone; Spring: 316SS; O-Ring: Fluoroelastomer, Other: Epoxy. Wetted materials This is an explosion-proof brass flow switch, actuation setpoint 0 50 GPM (1 89 LPM), and calibrated for water at standard conditions It is used for the accurate detection of excessive or insufficient flow rates...
B B u u r r n n H H a a z z a a r r d d W W a a r r n n i i n n g g s s (Yellow triangle with radiating lines) indicate a potential burn injury to personnel. (part S0001 for sink SS32), Fig 18 (part S0002 for sink SS64), and Fig 21 (part S0003 for sink SS96) on pages 10-12 (Yellow triangle with a lightning bolt) indicate a possible shock hazard is present.
To Operate the Soap Pump - press the soap spout down repeatedly to start the flow of the soap Fig 9 – Install Soap Spout Fig 10 – Secure Soap Spout with Bushing Fig 11 – Install Soap Dispenser Tubing Fig 12 – Close Up of Soap Pump with Tubing Attached www logiquip com...
11 Re-install the bushing onto the spout from inside the cabinet (Fig 10) as shown on page 10 12 Attach the soap tubing to the spout and soap pump (circled in red) as shown in Fig 11 and Fig 12 on page 10 www logiquip com...
SCRUB SINKS USE & INSTALLATION GUIDE PREVENTATIVE MAINTENANCE EVERY 6 MONTHS 1 Check and adjust the temperature setting EVERY 12 MONTHS 1 Shut off the water supply 2 Open up checkstops 3 Clean strainers and check for free movement of checkstop poppet 4 Replace seals if cracked, cut, or worn 5 Re-assemble 6 Adjust stem to the desired temperature...
Check Valve For Both Knee Operated or Infrared Activated Plumbing S0131 Sink Dividers (not shown) For Both Knee Operated or Infrared Activated Plumbing S0034 Sink Dividers (not shown) For Both Knee Operated or Infrared Activated Plumbing S0034 www logiquip com...
SCRUB SINKS USE & INSTALLATION GUIDE PLUMBING ACCESS ( MANUAL KNEE OPERATED SINKS ONLY) 1 Open the knee panel door by depressing the latch at the bottom of the panel door (Fig 24) 2 Swing the door fully open to access the hinges (Fig 25) 3 Push the hinge pins inward to release the door from the hinges (Fig 26) Infrared Controls (for Infrared Self-Activated Sinks only) 4 To re-attach the door, push in the hinge pins and fit the top corners of the knee panel door onto the left and right hinge leaves SOAP...
• Use only distilled water and a mild detergent applied with a clean, dry lint-free cloth to clean decals or printed labels • Cleaning agents can remove or smear any printing from decals and print labels • Cleaning agents can damage plastic materials used in manufacturing covers for electronic items such as infrared sensor face www logiquip com...
SCRUB SINKS USE & INSTALLATION GUIDE OPTIONAL IR TIMER ASSEMBLY An Electronic IR Timer Assembly (S0202-SST-KT, Fig 30) can be purchased and in- stalled on Infrared Activated surgical scrub sinks The installation would require some modification of the sink top in order to install INSTALLING OPTIONAL ELECTRONIC SINK TIMER DISPLAY 1 To mount the timer housing on the sink top, drill three (3) holes using the hole Fig 30 - IR Timer Assembly For IR Scrub Sinks...
Seconds, Minutes (_ _ _ _ _ _), Hours (_ _ _ __ _) , Hours: Minutes: Seconds NOTE: On initial start-up, as well as after any programming changes, it is necessary to reset the unit before beginning operation www logiquip com...
SCRUB SINKS USE & INSTALLATION GUIDE Specification Description TIMER (S0050) SPECIFICATIONS AND TERMINAL CONNECTIONS NPN, Contact Closure; Accumulates time when connected to common; Low State: < 1.0 VDC, High State: > 2.0 VDC Start/Stop Input (28VDC max) Specification Description The timer is an Elapsed Time Indicator which is powered by an internal 3 volt lithium battery It comes pre-assembled in its housing and Security Input Allows access to panel reset and programming features ready for mounting to the sink...
This drawing shows how the In-line Wall Carrier for 1-Station Scrub Sinks is assembled NOTE: To ensure stability, LogiQuip recommends that the In-line Wall Carrier Mount Assembly be tied into the existing wall support structure All applicable local and state building codes should be adhered to when mounting this unit...
Page 22
This drawing shows how the In-line Wall Carrier for 2-Station Scrub Sinks is assembled NOTE: To ensure stability, LogiQuip recommends that the In-line Wall Carrier Mount Assembly be tied into the existing wall support structure All applicable local and state building codes should be adhered to when mounting this unit...
Page 23
This drawing shows how the In-line Wall Carrier for 3-Station Scrub Sinks is assembled NOTE: To ensure stability, LogiQuip recommends that the In-line Wall Carrier Mount Assembly be tied into the existing wall support structure All applicable local and state building codes should be adhered to when mounting this unit...
Electronic Components, Cart Covers, Custom Products, LogiCell Tambour Doors, Label Holders, and Casters During this period, LogiQuip will, in its sole discretion, repair defects in materials or workmanship that existed when the product was received or replace the product Visit LogiQuip.com/Warranty...
Need help?
Do you have a question about the SS32 and is the answer not in the manual?
Questions and answers