Simv (Synchronized Intermittent Mandatory Ventilation) - O-Two e500 User Manual

Automatic transport ventilator
Hide thumbs Also See for e500:
Table of Contents

Advertisement

5.2.1. SIMV (Synchronized Intermittent Mandatory Ventilation).

In SIMV mode the ventilator will deliver volume ventilation at the
set Tidal Volume (Vt) and Rate (Figure 10).
The default trigger for this mode is 3 L/min. If trigger condition
is met, the ventilator will deliver synchronized volume controlled
mandatory ventilation.
In SIMV mode the selected breathing rate remains constant and
the time of spontaneous breathing window will change if patient
triggers the synchronized mandatory breath before the normal
start of inhalation phase (beginning of Ti).
If no effort was detected during the trigger period, the ventilator
will initiate mandatory ventilation at the end of trigger window.
Should the patient demand more flow than set by user, he/she can
withdraw the excess from ambient.
TABLE 2 - SIMV (Synchronized Intermittent Mandatory Ventilation)
PARAMETER
Tidal Volume/
Rate
I:E ratio
PEEP
Trig.
O
%
2
P max.
P min.
Manual
e500 User Manual
RANGE
100/20,150/20, 200/20, 250/20, 300/10,
350/10, 400/10, 450/10, 500/10, 550/10,
600/10, 650/10
0, 4, 5, 6, 7, 8, 9, 10, 12, 15, 18, 20
3 L/min
(100% or 60%)
10, 15, 20, 25, 30, 35, 40, 45, 50, 55, 60, 80
0, 2, 3, 4, 5, 6, 7, 8, 10, 12, 15, 20
Refer to Manual and I-Hold section
1:2
DEFAULT
500/10 or selected by
user during start-up
1:2
5
3
100%
30
3
ready
31