SERVICE MANUAL Analgesic Gas Mixing and Delivering System Electronic Nitrous Oxide/Oxygen Electronic Nitrous Oxide/Oxygen Analgesic Gas Mixing and Delivering System 01EQ3000 Made in Canada by O-Two™ Medical Technologies Inc. Part Number: 15PL1033 Rev. 3, July 2024...
• Never attach the face mask to the patient using a head harness. • The eAdvantage® shall not be used in oxygen rich nor fl ammable gases or anaesthetic agents environments. Keep away from open fl ames, sparks and grease/oil.
C) Pneuview Adult Test Lung set to 0.05 L/cmH O & Rp20. A cycling Test Fixture (O-Two e700 with Patient Circuit) it is used in the factory as a dual lung with triggering capabilities. Alternatively, other patient simulating devices could be used, as long as a steady rate and volume above 900 ml is maintained, such as a 1L syringe.
Power Input Connection Control Knob USB Connection Cancel Key Alarm Buzzer Audio Paused Key Green LED indicator for Oxygen Flush Key AC power Orange LED indicator for Oxygen Input Connection battery charging Airway pressure LCD Screen monitoring port eAdvantage Service Manual...
Check and test function of Sensors and Proportional Valves. 3.3 Main Operation A) To start the eAdvantage®, press the Control Knob (K) for 1 second, the associated green LED (J) will commence fl ashes at a high frequency. After 1 second, the eAdvantage® will turn ON and begin the calibration process.
Battery status will be displayed on the screen. There are two di erent status indicators, one showing discharge status and the other showing the charging status. Battery discharge status The battery discharge status indicator on screen should be around 50% of full eAdvantage Service Manual...
Disconnect the quick disconnect adapter on the N O supply line, sstart the stopwatch when pressure values have stabilized on the Timeter. The trapped pressures as indicated on the High Pressure Gauge must not drop more than 0.5 PSI in 30 sec. eAdvantage Service Manual...
O %) a) Having connected the supply hose to the regulators, ensure that the eAdvantage is in the OFF condition, then turn on the oxygen and nitrous oxide supply. Using a mild soap solution, spray the input connections to the eAdvantage to check for leaks. If any leak is detected, tighten the connection and re-test.
‘Warning’ symbol, while Medium and Low-priority alarms have a yellow symbol. The eAdvantage® device can display multiple alarm symbols simultaneously if several failures are occurring at the same time. In such cases, the audible alarm will correspond to the highest-priority alarm.
Page 11
≥ 80 PSI Solid symbol with a 1 Burst with 10 Empty High Directly fl ashing red warning pulses each, repeat Battery symbol every 7.5 seconds The solid yellow symbol with a fi xed Battery yellow warning symbol eAdvantage Service Manual...
5.2 Preventive Alarms The eAdvantage® will not switch ON if any of the problems listed below are detected: Valve Failure. O Valve Failure. Flow Sensor Failure. O Flow Sensor Failure. U12 (Paw) Sensor Failure. U13 (O input pressure) Sensor Failure.
5.3 Flow Valve Leak Alarm The eAdvantage® system includes a fl ow valve leak detection feature that functions when the unit is supplied with gas. While the device is ON, it continually monitors the outputs from both fl ow sensors.
O Flow Sensor O Input Connector Respiratory Pressure Sensor Input The Pressure Sensor Sensor Hose O Input The Pressure Sensor Removal Kit Inlet Manifold 14. Patient Circuit O Inlet Manifold 3-Bypass Valve Flow Control Valve O Flow Control Valve eAdvantage Service Manual...
Change the gas cylinder Battery discharges No proper charging/faulty Charge battery as per quickly battery instructions/replace battery Change battery/connect Battery empty/no power eAdvantage® cannot power supply/ send to be switched ON supply connected/defective O-Two™ for repair or service eAdvantage Service Manual...
01FM4999-CS Universal Face Mask Case/12 01CV0106* Power supply cord (Canada and the US) Each 01CV0104 External power supply Each 01CV7040-1 eAdvantage® Mounting Bracket w/ C-Clamp Each eAdvantage® Mounting Bracket 01CV7040-2 Each w/ Medirail clamp 01CV9200 Li-Ion Smart Battery Each 02RT1303...
Page 17
NITROUS OXIDE HOSES PART # DESCRIPTION UNIT O-Two™ 6 Foot (1.85 Meter) N O supply 01FV4303-AFN-N Hose with AFNOR Gas supply fi tting and Each N2O DISS nut Device Connection O-Two™ 6 Foot (1.85 Meter) N O supply 01FV4303-AGA-N Hose with AGA Gas supply fi tting and...
Yes (for Data Transfer & Software Update) Live Monitoring Alarms and setting only Parameter Settings Control Knob Lock Key Function Yes w/ passcode Dimensions (mm) 305 x 210 x 130 Weight With & W/O Battery (KG) 2.80 & 2.20 eAdvantage Service Manual...
Page 20
Your Representative is: O-TWO MEDICAL TECHNOLOGIES INC. For your nearest Authorized O-Two™ Distributor In North America call Toll Free 1-800-387-3405 MedNet EC-REP IIb GmbH SERIAL N Borkstrasse, 10 48163 Münster, Germany www.otwo.com 45A Armthorpe Road, Brampton, ON, Canada, L6T 5M4 Telephone: +1 905 792-OTWO (6896) N.A.