SAES SIP POWER User Manual page 34

Ion pump controller
Table of Contents

Advertisement

Bit 31:0 – Total working hours
These registers show the device total working hours (only the time spent supplying current).
TEMPERATURE – 0x3000 – Internal temperature
Bit
15
R/W
R
Bit
7
R/W
R
Bit 15:0 – Internal temperature
These registers contain the device internal temperature reading provided in Kelvin degrees.
ARCING_NUMBER – 0x3001 – Arcing events number
Bit
15
R/W
R
Bit
7
R/W
R
Bit 15:0 – Arcing events number
This register contains the number of arcing events since last start (or restart) command.
STATUS – 0x3002 – Power supply status
Bit
15
14
R/W
R
R
Bit
7
ALM_OVERTEMP ALM_INTERLOCK
R/W
R
Bit 15:13 – Reserved bits
These bits are reserved for future use.
Bit 12 – Communication alarm
This alarm latch is set when the device (Modbus Master or Ethernet Master) who started the power
supply (sent enable start or enable restart commands) does not send any request for KEEPALIVE
ms.
Bit 11 – Arcing alarm
This alarm latch is set when the power supply detects a arcing event.
Bit 10 – Output over-current alarm
This alarm latch is set when the power supply detects a output over-current event. This alarm latch
is set only if switch SW1 mode is set to ON and the output current increases over
M.HIST.0109.23 Rev.1
14
13
R
R
6
5
R
R
14
13
R
R
6
5
R
R
13
12
ALM_COMM
R
R
6
5
ALM_SAFE
R
R
USER MANUAL - SIP POWER ION PUMP CONTROLLER
12
11
TEMPERATURE[15:8]
R
R
4
3
TEMPERATURE[7:0]
R
R
12
11
ARCING NUMBER[15:8]
R
R
4
3
ARCING NUMBER[7:0]
R
R
11
ALM_ARCING
ALM_IOUT
R
4
3
ALARM
IOUT_GRADIENT
R
R
10
9
R
R
2
1
R
R
10
9
R
R
2
1
R
R
10
9
ALM_VOUT
ALM_VIN
R
R
2
1
NEED_RESTART
R
R
Page 34 of 62
8
R
0
R
8
R
0
R
8
R
0
ENABLE
R