7
AO
8
SW1 –
9
+5V
10
11
SW3 –
12
SW2 –
13
IP active –
14
IN –
15
SW1 +
If one of the above events reaches such a value the appropriate comparator-switch is connected
and the impedance between the terminals + and – is between 11 and 15 Ω. The maximal current
through the connected terminals must be limited up to 100 mA.
Maximal permissible voltage between switched off terminals + and – is 30 V.
Above-mentioned polarity for all terminals SW and IN must be kept. All pairs of terminals SW and
IN are galvanically insulated. The voltage between pairs must not exceed 100 V.
The supply works only when the control input INTERLOCK is active, that means, that a current
from 2 to 10 mA through the terminals IN+ and IN– must flow (see Sec. 3.2). Input impedance of
circuits IN is about 2 kΩ.
The IP ACTIVE comparator-switch is connected when the pump is powered (Active status).
For the output signals SW1 (over current alarm) SW2 and SW3 see Sec. 4.3
For the Analog output signal see next Sec. 8.2
8.2.

Analog Output

In the IN/OUT connector the ANALOG OUTPUT is provided for the current/pressure signal (pin 1
and 7). This signal is proportional to logarithm of ion pump current. The limits of the signal are: 0 V
for ≤1 nA; 10 V for 100 mA. Pressure scale, must be recalculated accordingly to ion pump
sensitivity and chosen pressure unit.
M.HIST.0109.23 Rev.1
analog output (to refer to ground)
negative terminal of switch 1 (over current alarm)
output of positive feeding voltage
not used
negative terminal of comparator switch 3
negative terminal of comparator switch 2
negative terminal of switch Operation IP on
negative terminal of control input INTERLOCK
positive terminal of switch 1 (over current alarm)
USER MANUAL - SIP POWER ION PUMP CONTROLLER
Page 28 of 62