SW3 Mode and SW3 Min Thr. and SW3 Max Thr.
SW3 is the same as SW2.
SW3 Mode is the mode (SIMPLE / WINDOW / OFF)
SW3 Min Thr. and SW3 Max Thr. are the thresholds.
IP Address, IP Netmask and MAC Address
IP Address (Internet Protocol Address), IP Netmask e MAC Address will show up in the menu only
if the instrument is equipped with the ETHERNET-USB card.
IP address can be changed according to user needs. It is a multifield parameter that allows to set
the 4 different bytes of an IP address.
IP Netmask field allow you to change Sip Power IP Netmask which is represented in CIDR
notation. For example the class 255.255.255.0 = 8+8+8+0 is indicated as 24
MAC Address of ETHERNET-USB card cannot be modified.
Keepalive
This parameter is a watch-dog on the remote communication. The parameter is the time (in
milliseconds) and indicates within how much time the instrument expects a communication from
remote connection.
It can be OFF (by setting it below 1 second), which means no control is performed. All other values
from 1000 to 900000 ms activate the watchdog after the START. This watchdog is related to the
communication channel through which the START has been received.
When there is no communication, the instrument resets to a security status: interrupts high voltage
emission and raises the "Communication" alarm.
Backlight Sleep
After these time duration the display back light is switched off to save energy. It can be modified
from 5 to 600 seconds or put to off so the backlight is always on.
Active Time
It indicates how long the SIP POWER supplied high voltage during its whole life.
Int. Temperature
Read-only internal temperature of SIP POWER.
Card Type
Readonly parameter that indicates which boards are installed in the instrument:
0 = base board with RS485
1 = base board with RS485 and Ethernet communication
2 = user interface with display and keys
3 = complete version with both user interface and Ethernet